Aer Use
• To deate, disconnect each of the air hoses
from the Camel. NOTE: The inial ‘rush’ of air as
the hoses are disconnected from the Camel, can
be reduced by rst disconnecng the air hose
from the Airo 24 and then pressing each of the
buons on the hand control.
• Roll up the Camel with the hand control in the
hollow (A) - g.
10
secure the straps and place
in the carrying bag - g.
11
Roune Cleaning
Fully inate all secons of the Camel, sponge clean
with a non abrasive proprietary liquid cleaner or
disinfectant and dry thoroughly. Do not dry using
hot air.
Disinfecng
The Camel should be fully inated and pre-cleaned
by washing with detergent and water, thoroughly
rinsed and then disinfected using a 1:10 soluon of
household liquid bleach and water (1 part bleach
to 10 parts cold water). Leave the soluon on
the Camel for approximately 10 minutes before
thoroughly rinsing o with clean cold water. Allow
to air dry before storage. Do not dry using hot air.
Maintenance
Other than roune cleaning, the Camel is
maintenance free.
We recommend that a thorough examinaon of
the Camel is carried out regularly; the frequency of
examinaon may need to be increased depending
on usage and operang environment.
If required, Mangar Health can provide the above
inspecon within an annual service contract. Please
contact our Customer Service Department for
further details (UK only).