Memory EraseMemory Erase
When the batteries are taken out of the monitor, all
records of readings saved in the memory will be
erased.
Note: When taking a measurement, the memory related
buttons are inactive.
Press the "Memory" button to enter the memory
function and select M1, M2 or M3, until you had
located the memory group you wish to erase, and
then press the "Mode" button and hold it down more
than 3 seconds until the "clr" appears on the display
which means you had successfully erased the
records of readings of the selected memory group.
9
To set Date and Time after proper battery installation
or replacement.
1. After the batteries have been properly installed,
"1 - 01" and "13:00" will appear as Date and
Time respectively.
2. Press the Mode button repeatedly, the individual
figure will start blinking sequentially from month, date,
hour to minute.
Mode
3.
After you have selected particular figure, press the
Recall/Set button to make adjustment.
Note : Pressing the button once will cause the
display to advance by one digit.
4. After you have done the adjustment, press the
Mode button or Start/Stop button to save your
settings .
Mode Start/Stop
When a valid measurement is made, the result
(systolic, diastolic, pulse rate, date and time) will be
stored in the memory automatically as soon as the
device is turned off, or when another measurement
is initiated.
The device is capable to store 3 x 12 readings,
and also computes averages of readings. When
more than 12 readings are recorded into memory,
the oldest ones will be deleted.
By pressing Memory button to select M1, M2 or
M3, "A" will be shown in the down left hand
corner, which means you have entered into
memory function. "A" means the average of all the
recorded readings. Now press the Recall/Set
button, then "A" will become "1", which is your
last reading recorded readings in the memory.
Press the button repeatedly, the records will be
displayed sequentially. Press the Start/Stop button
to exit the memory function.
Memory FunctionSetting Date and Time
Memory Recall
Memory Input
10
Memory Number
Error IndicatorsError Indicators
Err Indicators / Troubleshooting
Error indicator
Possible sources of errors
Correction:
Reapply the cuff and measure again.
Indication:
Could not obtain pulse rate.
Cause:
Extremely weak pulse or arrhythmia.
Correction:
Reapply the cuff and measure again.
Cause: Cuff get loose during the inflation.
Correction:
Reapply the cuff and measure again.
Cause: Cuff was loosely applied.
Incomplete inflation.
Indication:
Cause: Exhausted batteries or leaking cuff.
again.If same Err message indicated,
request a repair.
Correction:
Replace new batteries and measure
Indication:
Irregularly fast deflation.
during measurement.
Cause:
Excessive arm movement or talking
Cuff tighter, and then measure again.Correction:
Take 5 minutes break or reposition
Indication: Irregular pulse.
during measurement.
Cause:
Excessive arm movement or talking
measure again.
Correction:
Take 5 minutes break, and then