USER MANUAL
MI 404
Free & Total Chlorine
Meter
Dear Customer,
Thank you for choosing a Martini product. This
manual will provide you with the necessary infor-
mation for the correct use of the instrument.
Please read it carefully before using the meter.
This instrument is in compliance with directives.
Range 0.00 to 5.00 mg/L Cl
2
Resolution 0.01 mg/L (0.00-3.50 mg/L);
0.10 mg/L (above 3.50 mg/L)
Precision ±0.04 mg/L @ 1.50 mg/L
Light Source Tungsten lamp
Light Detector Silicon Photocell and 525 nm
narrow band interference filter
Method Adaptation of the USEPA
method 330.5 and Standard
Method 4500-Cl G.
Environment 0 to 50°C (32 to 122°F);
100% RH max.
Battery Type 1 x 9 volt
Auto-Shut off After 10' of non-use
Dimensions 192 x 104 x 52 mm
(7.5 x 4.1 x 2”)
Weight 380 g
SPECIFICATIONS:
MEASUREMENT PROCEDURE:
1• Turn the meter on by pressing ON/OFF.
2• When the LCD displays “- - -”,
the meter is ready.
3• Fill the cuvet with 10 mL of
unreacted sample, up to the
mark, and replace the cap.
4• Place the cuvet into the holder
and ensure that the notch on
the cap is positioned securely
into the groove.
5 Press ZERO and “SIP” will blink
on the display.
6• After a few seconds the dis-
play will show “-0.0-”. The
meter is now zeroed and
ready for measurement.
9• Immediately reinsert the cuvet into the
holder and ensure that the notch on the
cap is positioned securely into the groove.
10 Press READ and “SIP” will blink during
measurement.
11 The instrument directly displays concentra-
tion in mg/L of free chlorine.
12 Then, for total chlorine measurement, re-
move the cuvet and add to the reacted
sample 3 drops of DPD 3. Replace the cap
and swirl gently.
13 Reinsert the cuvet into the holder and en-
sure that the notch on the cap is posi-
tioned securely into the groove.
14 Wait for 2 minutes and
30 seconds then press
READ and “SIP” will
blink during measure-
ment.
15• The instrument directly displays concentra-
tion in mg/L of total chlorine.
Interferences
Bromine (positive error)
Chlorine dioxide (positive error)
Iodine (positive error)
Oxidized Manganese (positive error)
Chromium (positive error)
Ozone (positive error)
7• Remove the cuvet.
8• For free chlorine measurement: add 3
drops of DPD 1 indicator and 3 drops of
DPD 2 buffer to an empty cuvet.
Immediately add 10 mL of unreacted sample.
Replace the cap and swirl gently.
ISTMI404 03/05
Question and Answer IconNeed help?

Do you have a question about the Martini Instruments MI 404 and is the answer not in the manual?

Martini Instruments MI 404 Specifications

General IconGeneral
BrandMartini Instruments
ModelMI 404
CategoryMeasuring Instruments
LanguageEnglish

Summary

MI 404 Free & Total Chlorine Meter User Manual

Specifications

Details range, resolution, precision, light source, detector, method, environment, battery, auto-shut off, dimensions, and weight.

Measurement Procedure

Step-by-step guide for performing free and total chlorine measurements using the meter.

Instrument Operation and Maintenance

Display Codes Guide

Explains various codes displayed on the meter, including status, errors, and battery level indicators.

Error Messages

Details error messages, their meanings, and troubleshooting steps for zero and sample readings.

Battery Replacement

Instructions on how to safely replace the 9V battery in the meter.

Accessories and Warranty

Lists available accessories and provides warranty information for the meter.