MWC-E Series Ultraviolet Sterilizers
MWC-E7 / MWC-E10 / MWC-E20
Installation and Operation Manual
2022 VERSION

Questions and Answers

  • G
    garrett11Jul 29, 2025
    What to do if the alarm is chirping, the lamp is on, and the gland nut is glowing blue on my Master MWC-E Laboratory Equipment?
    • A
      Autumn ParkerJul 29, 2025
      If the alarm is chirping, the lamp is on, and the gland nut is glowing blue, it's likely that the 9000-hour timer has reached its limit because the lamp is a year old. To resolve this, replace the lamp and reset the 9000-hour timer.
  • M
    Mr. Frank ArnoldAug 12, 2025
    Why is the alarm chirping and the gland nut not glowing blue on my Master MWC-E?
    • C
      Charlotte LaneAug 12, 2025
      If the alarm is chirping and the gland nut isn't glowing blue, it could be due to a burned-out germicidal lamp or a defective power supply. You should replace the lamp and reset the 9000-hour timer, or replace the power supply.

Summary

Shipping Description

Pre-Treatment Requirements

Unit Location and Installation Zone

MWC-E SS Vessel Installation

Electrical Requirements and Power Supply Setup

Grounding Instructions

MWC-SERIES Grounding Lug Installation

Details on installing the grounding lug for MWC-Series UV systems to ensure a complete ground connection.

Component Installation Procedures

Quartz and Oring Installation

Step-by-step guide for correctly installing the quartz sleeve and o-ring into the UV vessel.

Germicidal Lamp Installation

Instructions for safely and properly installing the germicidal lamp into the quartz sleeve.

Power Supply Start Up

Procedure for initiating the power supply and confirming the germicidal lamp is operational.

Filling the UV Unit with Water

Steps for filling the UV vessel with water and purging air before operation.

Piping Disinfection and Final Checks

Disinfecting the Piping System with Chlorine

Protocol for disinfecting the water system using chlorine after UV unit installation.

Final Check Procedures

Essential checks to perform after installation and disinfection to ensure proper function.

Annual Maintenance and Timer Reset

Annual Germicidal Lamp Replacement and Quartz Cleaning

Procedure for yearly maintenance including lamp replacement and quartz sleeve cleaning.

9000 Hour Timer Reset Procedure

Instructions for resetting the 9000-hour timer after lamp replacement.

Troubleshooting MWC-E Series

4 Year Limited Warranty Information

Master MWC-E Series Specifications