OPERATING MANUAL:
REGLO DIGITAL PUMP
DRIVES WITH ADVANCED
CONNECTIVITY
MASTERFLEX
®
REGLO 78018-10
Model Nos.
78018-10
78018-12
78018-14
78018-20
78018-22
78018-24
78018-40
78018-42
A-1299-5201
Edition 01
(US & Canada only) Toll Free 1-800-MASTERFLEX • 1-800-637-3739
(Outside US & Canada) 1-847-549-7600 • 1-847-381-7050
www.masterex.comtechinfo@masterex.com
MASTERFLEX
®
REGLO 78018-40

Table of Contents

Questions and Answers

  • C
    Craig DavisAug 11, 2025
    How to troubleshoot motor not rotating and display not lighting on Masterflex Ismatec Reglo 78018-10?
    • D
      dwatsonAug 12, 2025
      If the motor of your Masterflex DC Drive doesn't rotate and the display doesn't light, it's likely due to a power issue. First, ensure the power cord is securely plugged into the drive and the mains power outlet, and that the power is switched on. Also, inspect the power cord for any visible damage and replace it if necessary.
  • D
    Dr. Michelle BrownAug 18, 2025
    What to do if Masterflex DC Drives pump doesn't rotate during remote operation?
    • P
      Patrick WeberAug 18, 2025
      If your Masterflex DC Drives pump doesn't rotate and the display doesn't light up during remote operation, the issue might be a defective remote control or a setting error. Here's what you can do: 1. Switch the pump drive off, then check that the DB-25 connection cable is fully inserted into the drive's port. 2. Switch the pump drive back on. 3. If the motor still doesn't rotate, select REMOTE from the Mode Selection screen and/or the Settings screen and verify that the settings are correct.
  • S
    Sarah LopezAug 27, 2025
    Why is the touchscreen unresponsive on my Masterflex Ismatec Reglo 78018-10 DC Drives?
    • C
      carriegillespieAug 27, 2025
      If the touchscreen on your Masterflex DC Drives is unresponsive, it might be because you are wearing thick latex gloves. Remove the gloves before using the touchscreen.
  • R
    Robert AdamsSep 3, 2025
    How to fix weak calibration strength on Masterflex DC Drives?
    • M
      mayleslieSep 3, 2025
      If you observe a weak calibration strength displayed on the calibration strength bar of your Masterflex DC Drives, it could be due to a poor calibration volume or flow rate. To improve calibration accuracy, consider using a larger calibration volume and/or altering the flow rate.

Masterflex Ismatec Reglo 78018-10 Specifications

General IconGeneral
Speed Range0.1 to 100 rpm
DisplayLCD
ReversibleYes
Communication InterfaceRS-232
Power Supply100 to 240 VAC, 50/60 Hz