5. PROBLEM SOLVING
If the "LOW BAT" icon is displayed on the LCD readout at any time,
the batteries should be replaced as quickly as possible.
When the unit is in the power on mode and the LCD displays "000%,"
the sensor is not connected properly. The "ATTACH SENSOR" icon
will also appear on the display. Check the sensor connection and if
the condition persists, contact Maxtec's Customer Service
Department.
If, at any time, "ErX" (i.e. Er1, Er4, etc.) appears on the LCD readout,
contact Maxtec's Customer Service Department.
6. CLEANING AND MAINTENANCE
When cleaning or disinfecting the MAXO2 Monitor, take appropriate
care to prevent any solution from entering the instrument.
The MAXO
2
Monitors surface may be cleaned using a mild
detergent and a moist cloth.
The MAXO
2
Monitor may be disinfected using standard topical
disinfectants.
The MAXO
2 Monitor is not intended for steam, ethylene oxide or
radiation sterilization.
Store the MAXO
2 Monitor in a temperature similar to its ambient
environment of daily use.
15
R213M05H.qxd 6/26/2006 9:04 AM Page 14

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Maxtec MAXO2 OM-25ME and is the answer not in the manual?

Maxtec MAXO2 OM-25ME Specifications

General IconGeneral
BrandMaxtec
ModelMAXO2 OM-25ME
CategoryOxygen Equipment
LanguageEnglish

Related product manuals