866.4.MAXTEC WWW.MAXTEC.COMiii
EXPLANATION OF ABBREVIATIONS
FIO2 Fractional Concentration of Inspired Oxygen
DISS Diameter Indexed Safety System
NIST Non-Interchangeable Screw Thread
psi Pounds Per Square Inch
l/min Liters Per Minute
SAFETY INFORMATION—WARNINGS AND CAUTIONS
DANGER
Indicates an imminently hazardous situation which, if not avoided, will result in
death or serious injury.
WARNING
Indicates a potentially hazardous situation which, if not avoided, could result in
death or serious injury.
CAUTION
Indicates a potentially hazardous situation which, if not avoided, may result in
minor or moderate injury.
CAUTION
Used without the safety alert symbol indicates a potentially hazardous situation
which, if not avoided, may result in property damage.
OR
CONSULT ACCOMPANYING DOCUMENTS
Symbol for “USE NO OIL
Symbol indicates the device complies with the requirements of Directive 93/42/
EEC concerning medical devices and all applicable International Standards.
Symbol for DO NOT USE
WARNINGS
Only trained, qualified medical personnel under the direct supervision of a licensed physician should
operate the Air-Oxygen Blender .
Use this Air-Oxygen Blender only for its Intended Use as described in this manual.
Confirm prescribed dose before administering to patient. Monitor on a frequent basis.
The Air-Oxygen Blender shall be serviced by a qualified service technician.
Always follow ANSI and CGA standards for Medical Gas Products, Flowmeters and Oxygen Handling.
The stand-alone gas mixer is not suitable for use with O2 93.
An Oxygen Analyzer/Monitor must be used to verify oxygen concentration.