EasyManua.ls Logo

Maxtec MicroMax R203P14 - Pre-Use Testing; Alarm Test

Maxtec MicroMax R203P14
114 pages
Print Icon
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
866.4.MAXTEC WWW.MAXTEC.COM4
PRE-USE TESTING
WARNING
Read this User Manual before installing or operating the Air-Oxygen Blender.
Confirm the concentration of air/oxygen with an Oxygen Analyzer/Monitor.
CAUTION
Inspect the Air-Oxygen Blender for visual damage before use,
DO NOT USE if damaged.
NOTE: The tests listed below should be performed prior to placing the Blender in service.
Pre-Use Testing consists of:
Alarm Test
Reverse Gas Flow Procedure
1. Secure the Air-Oxygen Blender to a wall or pole bracket in an upright position.
2. It is recommended to install a condensation trap in the air supply line.
3. Connect the air and oxygen supply lines to the appropriate inlet fittings on the bottom of the
Blender.
4. Attach a flowmeter, or other metering device to one of the outlet ports and verify FIO2 range
for accuracy with an oxygen analyzer.
Primary Outlets Flow capacity:
High Flow Blender (R203P14 Model) 15 l/min to 120 l/min
Low Flow Blender (
R203P13 Model
) 3 l/min to 30 l/min
Auxiliary Outlet:
The auxiliary flow outlet maintains the same flow capacity and FIO2 accuracy as the Primary Outlets
with Bleed Valve not engaged. When bleed flow is activated, some of the air/oxygen mixture will vent
to atmosphere to maintain FIO2 concentration accuracy at the Low Flow settings.
High Flow Blender (R203P14 Model) 15 l/min or less
Low Flow Blender (R203P13 Model) 3 l/min or less
5. Attach a supply line to the outlet port of the flowmeter.
ALARM TEST
1. Connect the Air-Oxygen Blender to air and oxygen sources, pressurize the Blender and turn
“ON” the flowmeter.
2. Set Oxygen Concentration Dial to 60% FIO2.
3. Disconnect or turn “OFF” the air supply to the Air-Oxygen Blender. The Blender should alarm
with a loud whistle noise. The whistle indicates the alarm is operating correctly.
4. Reconnect and activate the air supply line to the Blender, the alarm should stop whistling.