866.4.MAXTEC WWW.MAXTEC.COM1
Transport / Storage Requirements
TEMPERATURE RANGE: -10°F to 140°F (-23°C to 60°C)
HUMIDITY: Max 95% Noncondensing
FIO
2 ACCURACY:* ± 3% of full scale
PRESSURE DROP:
Low Flow:
2 psi (0.14 bar) at inlet pressures from 30-90 psi (2.1- 6.2 bar) and at
10 l/min flow rate at 60% FIO
2.
High Flow:
3 psi (0.21 bar) at inlet pressures from 30-90 psi (2.1- 6.2 bar) and at
30 l/min flow rate at 60% FIO
2.
The Air-Oxygen Blender has been cleaned for Oxygen Service prior to delivery.
The Air-Oxygen Blender reverse gas flow complies with clause 6 of ISO 11195.
The Oxygen Analyzer should comply with ISO 80601-2-55.
DRYNESS AND COMPOSITION FOR INLET GASES:
Air: Medical Air supply should meet the requirements of ANSI Z86.1 - 1973
commodity specification for Air, type 1 grade D or better.
Oxygen: Oxygen supply must meet all requirements of USP Medical Oxygen
Grade N.
Dew Point: Both inlets should remain 10°F (5.55°C) or more below the lowest
temperature to which the air distribution system equipment is
exposed. At a temperature of 25°F (-3.9°C) and a pressure of
90 psi (6.33 kg/cm2) this equates to 2000 mg/m3.
* Accuracy of oxygen concentration will be affected if bleed is not activated at flow settings below 15 l/min for
the high flow Blender, and 3 l/min for the low flow Blender.
* Specifications are subject to change without prior notice.