Operating Manual
Operation instructions Page16-2
16
.1
Components
of
Third-Party Manufacturers
MBRAUN Glove Box systems are partly equipped with third-party manufacturers’
components such as:
•
Vacuum pump(s)
•
Compressor(s) for the system’s compressor cooling
•
Compressor(s) for refrigerator systems
•
PLC control components (Siemens)
•
TOUCH Screen Operation Panel (Siemens)
The original third-party manufacturers’ documents, in which the maintenance and
service of the components are described, are included in the systems delivery.
16.2 Regular Maintenance and Service
Main glove box and window Clean the exterior using
conventional detergents (do not use
caustic detergents); for this purpose
use a soft, lint free cloth; or a
vacuum cleaner if available, using a
brush attachment.
Note: If the box is equipped with an MBRAUN Clean-Jet unit then the interior of the box
and window may also be vacuumed with a brush attachment.
Gloves Check the gloves for damage; in
addition, use linen gloves to avoid
humidity in the box gloves.
Cau
t
ion
:Do not use powder within the box or within a clean room environment.
Replace gloves when damaged
- by no means attempt to repair gloves.