Operating Manual Page 3-3
If the
MBRAUN system is only part of a larger complete unit, the requirements of the other
system components have to observed as well.
3.1.3 Installation and Commissioning
We recommend, that a
MBRAUN technician installs the system and performs the first
commissioning. If the customer does the installation and first commissioning by himself, please
adhere to the following instructions.
3.1.4 Final Inspection
A final inspection and specification test is performed by a
MBRAUN technician if the installation is
made by
MBRAUN. If the customer does the installation and first commissioning by himself,
MBRAUN recommends to perform a complete system and specification test prior to operating the
system.
3.2 Gases Used by the System
3.2.1 Overview
Pressure
Reducing Valve
to customers
from gas
Working Gas IN
Regeneration Gas
V
acuum-
-
exhaust system
source 200 bar 6 bar (0.6 MPa)
Exhaust
depressurized
permanent)
(20 MPa)
EASY
LAB
from gas
Regeneration Gas IN
source 200 bar 0.3-0.4 bar (30 – 40 kPa)
permanent)
(20 MPa)
KF40
40
KF40
pump
5-6.0 bar (0.5 – 0.6 MPa)
Purge Gas IN
to customers
exhaust system
Purge gas OUT
GLOVE BOX