3.4.4 Connecting the Disposal Piping for Vacuum Waste Gases
► Connect the disposal piping between the vacuum pump exhaust and the customer’s
disposal facility (exhaust). Follow the manufacturer’s instructions for the vacuum pump
connections.
Connection must be depressurized.
!
Depending on the place where the vacuum pump is used an oil mist
filter can be used instead of the disposal piping. Contact
MBRAUN
Service for information.
3.4.5 Power Connection
The connection needs to be made to protected (fused) power supply that is equipped with a
CPC (earth conductor). The required values for connection: see the type plate. For Easylab
the power supply is 1 ph 220V AC 16A.
DANGER
Hazardous electrical voltage!
Risk of electric shock on direct and indirect contact with live parts with
the possibility, involuntary muscle reaction, muscle paralysis, burnt
tissues and organs, or death.
► All works on the electrics is only allowed to be performed by an
electrician. This includes the connection to the main power
supply.