Operating Manual
Operation instructions Page 4-3
4.2.3 Connect regeneration gas OUTLET
1
Connect “Regeneration gas OUTLET” with exhausting device or port which provide
by customer
2
Exhausting pipe line inside must be negative pressure
4.2.4 Connect vacuum pump OUTLET
1
Use piping to connect vacuum pump OUTLET with customer exhausting system.
Refer to manufacturer vacuum pump connection instruction
2
Exhausting pipe line inside must be setting negative pressure
Notice:
According to vacuum pump different using situation, you can use oil fog filter to instead of
exhausting system. Details please refer to service@mbraunchina.com
4.2.5 Connect main switch
Main power connection must be equipped with ground protection. Desired voltage is marked
on the nameplate of equipment.
Notice:
Setting pressure correctly as required.
Too high pressure will cause system damage, too low pressure will cause system unable to
run
Notice:
When regeneration gas exhausts, it will produce terrible gas. It is no pollution to the
environment and no harm to body. However, regeneration gas shouldn’t be exhausted
directly to the room inside. Manufacturer isn’t responsible for this behavior.
When use polluted or radioactive material, gas of the box cann’t be exhausted to room
inside.