Operating Manual
Operation instructions Page9-2
9.1 Profile
MBRAUN Glove Box remove water an oxygen
through gas cycle, that means the working gas
cycle in box and purification column constantly .This
way can guarantee system stability, economic
effectively to purify gas.
Notice:
The circulation mode must be keep opening
when Glovebox is used .Only in this way can
make the box gas purification constantly to keep
water/oxygen levels remain <1ppm.
Circulation mode can operate through control panel
button and indicate working state through the light.
Dia.1
Oxygen/moisture levels will rising when purification column is used after a period of time. So
purification column should regeneration regularly, at least when seeing oxygen/moisture
content rise up then active the regeneration process. Please refer to “regeneration section”.
While doing regeneration, the circulation should be turned off.
9.2 Necessary condition
Read all of the above chapters already.
All of the interface is connected well
All doors are closed
Box has been cleaned
System has been turned on
Purification column without regeneration state.
9.3 Circulation mode