Operating Manual
Operation instructions Page10-2
10.1 General information
After the purifier working a period of time, Oxygen/moisture is saturated in purifier. Now in
order to recover the ability on removing moisture and oxygen, it should perform regeneration
program.
Regeneration regularly, especially the content of water and oxygen rising .Regeneration of
interval depend on the status and using of customers.
Suggestion:
10.2 Necessary condition
Read all of the above chapters already
All of the interface is connected well
Enough regeneration gas
System has been turned on
Loop closed
10.3 The regeneration program
10.3.1 Active regeneration program
Attention;
Make sure there is enough regeneration gas and working gas before start reg. program and
check the flow when click regeneration button the first time.
Confirm the regeneration cycle by following methods: Do the regeneration when the
content of moisture and oxygen can’t remain below 1ppm and record running time,
also this time minus 10 hours as the time of the next regeneration.