Operation instructions Page12-5
Close hand valve ( DN10)
12.4.4 Take material out of antechamber
·Open inner door
·If a sliding tray is available: Pull out sliding tray; take out material from the tray; then slide
the tray into antechamber.
·If no sliding tray is available: Transfer the material directly into box
·Close inner door
12
.5 Transferring Material out of the Box
12.5.1 Preparation
► Observe Item “Important Notes” in this chapter.
► The antechamber door located outside the box is closed.
► Open the inner antechamber door.
► If a sliding tray is available: Pull out sliding tray; lay material on tray; then slide the tray
together with the material into antechamber.
►If no sliding tray is available: Transfer the material directly into antechamber.
►Then close inner antechamber door.
12.5.2 Take material out of antechamber
·Open outer door
Suggestion:
The antechamber valves must be returned to closed-position after each step of the
antechamber operation.
Both the inner door & outer door should keep closed when the antechamber is not in using.
After opening the outer door, the antechamber should be evacuated at least once, to prevent
the atmosphere of the glovebox.