Operating Manual
Operation instructions Page13-11
DANGER
Risk of injury!
The sensor element is continuously exposed to the box gas and
therefore is at risk from contamination.
► Ensure that all local and national safety guidelines are followed
when handling potentially contaminated materials.
!
NOTE!
MBRAUN recommends a maintenance cleaning procedure every 3
months.
CAUTION
Risk of damage!
When cleaning the sensors it is important that contamination from
the ambient atmosphere is prevented. Therefore, MBRAUN
recommends that the box parameters are set to a pressure of
between +1.0 and +5.0 mbar (see parameters chapter) and that the
circulation mode is switched OFF.For Glove Box systems that must
remain at a negative pressure, special procedures may be required.
Contact
MBRAUN service for advice.
This routine maintenance consists in cleaning the platinum winding of the MB MO-SE-1 and
moistening it with phosphorous acid H3PO4. The following aids are required for
disassembling and maintaining the MB MO-SE-1.
Tool for disassembly (screwdriver)
Soft, absorbent, lint free cloth (cotton)
Small quantity of phosphorous acid (H3PO4).
Protective clothing, including gloves and goggles
One dummy plug for the open circulation piping (DN40)