EasyManua.ls Logo

MCP CQ5010B - Page 12

MCP CQ5010B
13 pages
Print Icon
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
OS-10V 02/04/2001 Rev.2
11
For rise time and fall time measurements, the 10% and 90% amplitude points are used as
starting and ending reference points.
(1) Apply a signal to the input jack [18].
(2) Use the VOLTS/DIV and VAR controls to adjust the waveform peak to peak height to
five divisions.
(3) Adjust vertical position so that the tops of the waveform coincide with 100% point,
while the bottoms of the waveform coincide with 0% point.
(4) Adjust Sweep switch to obtain the positive-going direction or negative-going
direction of the waveform on the screen.
(5) Use the horizontal POSITION control to adjust the 10% points to coincide with a
vertical reference line.
(6) Measure the horizontal distance in divisions between the 10% and 90% points on the
waveform (divisions).
(7) Pulse rise time and fall time measurement is calculated by the following equation:
Horizontal distance (div) × sensitivity (Time/div)
Rise Time = -----------------------------------------------------------
Horizontal factor
For the example shown in Fig.4-7, the horizontal distance from 10% to 90% is 2.4
divisions, the sweep TIME/DIV setting is 1µS/DIV, factor x 1. The rise time is
calculated as follows:
1µS/DIV × 2.4DIV
Rise Time = ------------------------------ = 2.4µS
1
Fig. 4-7
4.4 TV Signals measurement
Steps :
(1) Connect TV signals to INPUT jack [19]
(2) Set Trigger method to “TV” [10], Sweep switch turn to 2mS/Div.
(3) Observe the screen, it should be negative synchronize pulse wave.
(4) Adjust VOLTS/DIV and VAR to obtain proper range.
4.5 X– Y mode applications
10%
90%
100%
2.4DIN