25
Guidance and Manufacturer’s Declaration - Electromagnetic Immunity
The RP Check is intended for use in the electromagnetic environment specified below. The customer
or the user should assure that it is used in such an environment.
Immunity Test
IEC 60601 Test level
Compliance
level
Electromagnetic environment
guidance
Conducted RF
IEC 61000-4-6
Radiated RF
IEC 61000-4-3
3 Vrms 150 KHz to 80
MHz
3 V/m 80 MHz to 2.5
GHz
[V1] V
[E1] V/m
Portable and mobile RF
communications equipment should be
used no closer to any part of the RP
Check, including cables, than the
recommended separation distance
calculated from the equation applicable
to the frequency of the transmitter.
Recommended separation
distance
d = [ 3.5 ] √ P
V1
d = [3.5] √ P 80 MHz to 800 MHz
E1
d = [ 7] √ P 800 MHz to 2.5 GHz
……E1
Where P is the maximum output power
rating of the transmitter in watts (W)
according to the transmitter
manufacturer and d is the
recommended separation distance in
metres (m).
Field strengths from fixed RF
transmitters, as determined by an
electromagnetic site survey, a, should
be less than the compliance level is
each frequency range. b
Interference may occur in the vicinity of
equipment marketed with the following
symbol:
NOTE 1 At 80 MHz and 800 MHz, the higher frequency range applies.