What to do if MEDAP MEDAP-BORA UP 2080 Medical Equipment has reduced or no flow rate?
- JJeremiah JohnsonAug 14, 2025
If your MEDAP Medical Equipment aspirator shows a reduced or no flow rate, make sure the septic fluid jar cap is correctly positioned. Also, check if the hydrophobic filter is clogged (if the vacuum gauge indicates vacuum, replace the filter). Inspect the tube for cracks and replace it if necessary. Replace any contaminated or porous seals. Ensure the septic fluid jar cap closes properly, replacing it if the clamp is bent. If the septic fluid jar is full, empty it and clean or replace the jar and mechanical overflow protection. Clean the overflow protection device or replace the septic fluid jar cap if it's contaminated with exudate. Finally, check and clean the tube connection in the septic fluid jar cap if it's clogged.