OPERATING INSTRUCTIONS
SURGICAL ASPIRATOR
MEDAP-TWISTA SP 1070
GA 5752 1555 DE 28

Table of Contents

Questions and Answers

  • T
    Tracey PowersAug 16, 2025
    What to do if MEDAP WISTA SP 1070 aspirator does not start and the display is off?
    • R
      Raymond CantuAug 16, 2025
      If your MEDAP Medical Equipment aspirator doesn't start and the operating status display is not illuminated, consider these possibilities: * The aspirator might be in standby mode. Try switching off standby mode using the foot switch. Operating the foot switch again will return the aspirator to standby mode. * The equipment or mains plug may not be properly seated in the socket. Check to ensure both are making proper contact. * There may be none or improper voltage. Verify the building fuse and check the specifications on the type plate. * The mains fuse could be defective. If so, replace the mains fuse.
  • S
    Scott HigginsAug 20, 2025
    How to fix MEDAP WISTA SP 1070 aspirator when it doesn't start but the display is on?
    • M
      mritterAug 20, 2025
      If your MEDAP Medical Equipment aspirator's operating status display is illuminated, but the aspirator does not start, it could be due to the following reasons: * Vacuum may still be present. To resolve this, switch off the aspirator, turn the regulating knob to the left, and then switch the aspirator back on. * The motor may be defective. In this case, have the equipment repaired by a service technician authorized by ATMOS.
  • P
    Phillip HernandezAug 23, 2025
    Why MEDAP Medical Equipment aspirator working but vacuum gauge indicates no vacuum?
    • K
      keith66Aug 25, 2025
      If the MEDAP Medical Equipment aspirator is working, but the vacuum gauge indicates no vacuum, the vacuum gauge is defective. Have the equipment repaired by a service technician authorised by ATMOS.
  • D
    Debra PetersenAug 27, 2025
    How to resolve vacuum regulation issues in MEDAP WISTA SP 1070 Medical Equipment?
    • K
      Kevin WallAug 27, 2025
      If the vacuum cannot be regulated on your MEDAP Medical Equipment, the membrane regulator is defective. Have the equipment repaired by a service technician authorised by ATMOS.
  • K
    Kayla GarzaAug 31, 2025
    What to do if MEDAP WISTA SP 1070 aspirator has been exposed to oversuction?
    • D
      Denise RichardAug 31, 2025
      If the MEDAP Medical Equipment aspirator has been exposed to oversuction, and there is no mechanical overflow protection device and no hydrophobic bacterial filter installed, the aspirator may no longer be used. Have the equipment repaired by a service technician authorised by ATMOS.
  • D
    dalvaradoSep 4, 2025
    What to do if MEDAP Medical Equipment aspirator operates but the display is not illuminated?
    • B
      Brittany ClarkSep 4, 2025
      If the MEDAP Medical Equipment aspirator is operating, but the operating status display is not illuminated, the LED on the operating status display is defective. Have the equipment repaired by a service technician authorised by ATMOS.
  • E
    esantosSep 7, 2025
    Why MEDAP Medical Equipment cannot be switched on and off?
    • J
      Jessica Robinson PhDSep 7, 2025
      If your MEDAP Medical Equipment cannot be switched on or off, the electronics are defective. Have the equipment repaired by a service technician authorised by ATMOS.
  • A
    Albert SmithNov 11, 2025
    What to do if MEDAP WISTA SP 1070 aspirator has reduced flow rate?
    • K
      kathrynrileyNov 11, 2025
      If your MEDAP Medical Equipment aspirator has a reduced or no flow rate, here are several possible causes and solutions: * The septic fluid jar cap may not be in the correct position. Ensure it is properly positioned. * The hydrophobic filter could be clogged. If the vacuum gauge indicates vacuum, replace the hydrophobic filter. * There might be a crack in the tube. Replace the tube. * The seal could be contaminated or porous. Replace the seal. * The clamp may be bent, preventing the septic fluid jar cap from closing. Replace the septic fluid jar cap. * The septic fluid jar may be full, causing the mechanical overflow protection to close. If the vacuum gauge indicates vacuum, empty the septic fluid jar and clean or replace the septic fluid jar and mechanical overflow protectio...