This service manual is for the information of service personnel only. Service may only
be conducted by personnel who
Has basic knowledge of electronic circuit and mechanical.
Has basic knowledge of medical devices and clinic.
Has received proper training of maintenance service of MEDCAPTAIN and has
certain knowledge of device principle, structure, performance and operation.
Has the necessary equipment and instruments.
The permission of MEDCAPTAIN.
1.2 The Limitations of the Service Manual
This service manual describes all performance and configurations of the device. Some
other devices performance and configurations are not included here. There are
differences between the published manual and the actual status of the device, especially
when device has been modified, a corresponding maintenance information is needed.
Therefore, this service manual may be used together with some follow-up complementary
information.
MEDCAPTAIN will complement the relevant information timely according to the actual
situation of the device modification.
Service personnel should also refer to the Operation Manual.
1.3 Device Maintenance and Inspection
The power supply of this device can reach to 100-240V. Maintenance against the
requirements of the service manual may cause an electric shock hurt or death.
Maintenance against these requirements of this service manual may seriously
damage the device.
The service personnel must be trained and permitted by MEDCAPTAIN.
The maintenance must be of electrostatic discharge (ESD) protection conditions. Do
not touch PCBA or semiconductors by hand without any protective measure.
Do not operate on the display by using sharp objects. Otherwise, the display may be
damaged.
Do not disinfect the infusion pump by using the high-pressure steam sterilization
method.
Before internal battery operation, check the battery to ensure that sufficient power is
available. Recharge, if required.
Liquid intrusion into the AC power socket, USB or nurse call socket may cause
short-circuit. While connecting the power cable, check if the connecting parts are dry.