Cleaning and Disinfection
28/ 46
Be careful not to scratch the LED light and camera during the cleaning. In
addition, do not use a hard object to scrub the transparent baffle in front of the
camera. Otherwise, scratches may be caused, which will affect the image
display.
8.3 Disinfection
1. Before the disinfection, clean the device according to the method provided in
section 8.2.
2. Completely immerse a piece of soft medical gauze in the recommended
disinfectant solution, wring out the gauze, and then use the gauze to wipe the
device surface.
3. Wipe all the surfaces of the device in sequence. For the contact time of the
disinfectant, see the operation manual of the disinfectant.
4. Ensure that all the edges and corners of the device are completely disinfected.
5. After the disinfection, immerse another piece of soft medical gauze in clean
water, wring out the gauze, and then use the gauze to wipe the device surface
for removing the residual disinfectant solution.
The following table lists the disinfectants recommended for the device and
the required contact time for the disinfection.
Table 8-2 Recommended disinfectant solution
Disinfectant Solution Name