SONNET 33
Warning Problem Solution
for max. 5 min.
The DL-Coil and implant are
disconnected.
Position the DL-Coil over the implant.
SONNET is in microphone
monitoring mode.
Switch SONNET off and on again.
The DL-Coil is positioned over
the wrong implant.
Position the DL-Coil over the correct
implant.
The DL-Coil cable is defective.
If the DL-Coil cable fails, the
DL-Coil will automatically power
off after 5 minutes.
1. Replace the DL-Coil cable.
2. Reposition the DL-Coil over the implant.
3. Switch your SONNET off.
4. Switch your SONNET back on.
10 seconds
The DL-Coil has powered off. 1. Reposition the DL-Coil over the implant.
2. Switch your SONNET off.
3. Switch your SONNET back on.
Random signals
The DL-Coil cable is defective. Replace the DL-Coil cable.
0 1 2 3 seconds