Medline ReadyBath MSCWARMER 14S User Manual

Medline ReadyBath MSCWARMER 14S
2 pages
To Next Page IconTo Next Page
To Next Page IconTo Next Page
Page #1 background imageLoading...
Page #1 background image
Hit
ONCE
Use arrows to
increase or decrease
temperature
RECOMMONDED
FACTORY
DEFAULT
125°F
TO ENTER SET-POINT TEMPERATURE
See
Unit
PAGE 1
Power Source: AC 115 (+/-10%), 60HZ, Max Watts: 300 (14S and 28S)
Maximum Temperature: 54
o
C / 130
o
F
Operating Environment: Indoor use, altitude up to 6,500 ft (2,000m).
Installation Category II, Pollution Degree 2,
ambient Conditions 10
o
C / 50
o
F to 35
o
C /95
o
F, 80% RH maximum.
Storage Temperature: -10
o
C / 14
o
F to 70
o
C / 158
o
F, 70% RH maximum
Approvals: Underwriter's UL / CUL, United States / Canadian for laboratory equipment.
Compliance: UL Standard 61010-1, IEC 61010-1 .
Exterior: Powder Coated Steel Interior: Aluminum
Insulation: Fiberglass Door: Acrylic
Thermo-control: PID Microprocessor Heater: Resistive-Tubular Incoloy
Unit Construction
MHSe WARMER14S MHSe WARMER28S
Capacity: 14 ReadyBaths Capacity: 28 ReadyBaths
Exterior Dimensions: 12.87"(32.6 cm)W x Exterior Dimensions: 22.14" (56.2cm)W x
18.00" (45.7 cm)H x 13.43" (34.1 cm)D 18.00" (45.7 cm)H x 13.25" (33.6cm)D
Shipping Weight: 29 pounds / 13.1 kg Shipping Weight: 34 pounds / 15.4 kg
Net Weight: 23.5 pounds / 10.6 kg Net Weight: 28.5 pounds / 12.9 kg
Unit Specifications
The control's self diagnostics are part of Intelligent Warmer's "watch-dog" system that continuously monitor the warmer's
operating system, temperature and performance. In the event of any of the operating alarm conditions, an alarm prompt
will continuously appear on the control display. The alarm conditions are described below. All alarm conditions will
automatically discontinue heater operation.
Medline Industries, Inc. Three Lakes Drive Northfield IL 60093 1-800-MEDLINE
Control prompts will only display when a fault or alarm condition exists.
Control Self Diagnostics
Indicates Connection Error
ALARM Codes "Err1" & "Err6"
or
Check the wiring connections
Indicates Open Input / No Sensor Error
ALARM Code "- - - -"
Check connection between Controller
and Thermocouple
Indicates Internal Memory Error
ALARM Code "E111"
Turn Controller OFF and On. If problem
remains, replace Controller
Indicates Watchdog Error
ALARM Code "Err4"
Controller needs to be replaced
Indicates Thermocouple Sensor Error
ALARM Code "Err8"
Turn off and wait until ambient
temperature recovers
Indicates Sensor Temp Over/Under Error
ALARM Code "nnnn" & "vvvv"
Increase or reduce process temperature
to get back in range
or
Operating Specifications
Issue Rev C 10 19
Controller Functions Buttons
MENU
Functions
Decrease
Increase





The warmers’ controller allows you to set and change the
units operating temperature between 100°F and 130°F,
and change the read out units to either °C or °F. These are
the only settings that can be changed. There are other
controller settings that are visible, but for safety and opera-
tional purposes can not be changed.
Hit
TWICE
Use arrows to
change temperature
units readout
FACTORY
DEFAULT
°F
TO CHANGE TEMPERATURE READ OUT
See
Unit
For questions regarding:
Technical Support
General Product Information, please contact:
Technical Support/Parts Order, please contact: PH: 1-800-482-HEAT FAX 773-622-2282
Copyright Quincy Lab Inc. 2018
Warmer Maintenance
Periodic cleaning is the only required maintenance for the Warmer. Disconnect the power cord before cleaning. To
clean interior and exterior surfaces, use a soft damp cloth. Avoid excessive amounts of water that could drip into the lower
plenum. Use a mild detergent or cleaner if necessary.
The door should only be cleaned using a lint-free cloth. Paper towels can mar the surface of the door. Do Not use any
commercial or alcohol-based cleansers on the door. This will cause cracking of the acrylic over time.
Medline Industries, Inc. Three Lakes Drive,
Northfield, IL 60093
1-800-MEDLINE (633-5463)
medline.com | info@medline.com
PAGE 4
To return to temperature display, aer making any change, simply press the buon unl you see the warmer’s temperature
on the display. If you are in an area of the MENU not shown press and hold unl you see the warmer’s current temperature.
P
P
Models MSCWARMER 14S & 28S
Operating Manual
NEXTGEN Warmer
INTELLIGENT WARMER
125
P
F
WARMING - Solid Red LED - Pack has not yet reached required temperature
HEATED - Solid Green LED - Packs have reached the required temperature and are ready to use.
TAKE FIRST - Green LED Flashes - Pack has been in the warmer for 84 hours. These packs should be used first.
DISCARD - Red LED Flashes - Pack has been in the warmer longer than 250 hours and should be discared.
INTELLIGENT WARMER
125
P
F
WARMING - Solid Red LED - Pack has not yet reached required temperature
HEATED - Solid Green LED - Packs have reached the required temperature and are ready to use.
TAKE FIRST - Green LED Flashes - Pack has been in the warmer for 84 hours. These packs should be used first.
DISCARD - Red LED Flashes - Pack has been in the warmer longer than 250 hours and should be discared.
Model 28S Model 14S
Question and Answer IconNeed help?

Do you have a question about the Medline ReadyBath MSCWARMER 14S and is the answer not in the manual?

Medline ReadyBath MSCWARMER 14S Specifications

General IconGeneral
CategoryFood Warmer
Power SourceElectric
Temperature ControlAdjustable
CertificationUL Listed
Product NameReadyBath MSCWARMER 14S
MaterialPlastic

Summary

Controller Functions and Operation

Control Panel Buttons Overview

Explains the primary buttons for setting temperature and read-out units on the control panel.

Temperature Adjustment and Readout

Guides on entering set-point temperature and changing temperature read-out units (°F/°C).

Control Self Diagnostics and Alarms

Details alarm codes, their meanings, and how the control displays fault or alarm conditions.

Warmer Maintenance and Support

Provides guidelines for cleaning the warmer and contact information for technical support.

Safety Precautions and Setup

Important Safety Warnings

Highlights crucial safety information regarding electrical shock and fire risks during operation.

Setup and Installation Guidelines

Instructions for proper placement, ventilation, and power connection for the warmer unit.

General Operation Procedures

Basic steps for powering on the unit and understanding its initial heating and temperature holding.

Safety Alarms and Indicators

Electronic Fail Safety Alarm Reset

Procedure for resetting the unit when the 'ALM' light indicates an over-temperature condition.

Thermal Safety Alarm Reset

Instructions for resetting the thermal safety shut-off triggered by electronic failure.

Indicator Sensor Status Explained

Description of LED light sequences indicating product warming, heating, and discard status.

Product Loading and Tips

ReadyBath Product Loading

Guidance on how to insert ReadyBath products into the warmer slots for proper warming.

Product Loading Best Practices

Tips for loading products to ensure quality and prevent staining of the aluminum shelves.