Control Unit Functions
Max. Patient Weight: 250kg
Power:
Ensure the rocker switch at the side of the control unit is in the ON
position.
Press the POWER/MUTE/UNLOCK button to switch the control
unit on.
A green LED will light up when the control unit is in use.
For clinical and technical support 24/7/365, call Medstrom:
UK: 0845 371 1717
IRE: 01 686 9487
DO NOT REMOVE FROM MATTRESS
!
SM547_REV3_0121
Quick User Guide: Medstrom Aria PRO
Unlock:
The control unit will automatically lock out after 3 minutes of
inactivity.
Press and hold the POWER/MUTE/UNLOCK button for 3
seconds to unlock the control panel.
Constant Immersion:
The control unit will default to Constant Immersion mode when
switched on.
Constant immersion is a reactive therapy in which all air cells
maintain constant pressure to maximise contact area and
redistribute pressure.
A green LED will illuminate on this button when Constant
Immersion is active.
Gradient Immersion:
Gradient Immersion (or ALP therapy) is an active therapy in
which air cells will alternate to provide o-loading.
This mode can be selected by pressing this button.
The cell cycle options of 5, 10 or 15 minutes can also be
changed by pressing this button.
The chosen cycle time will be visible on the digital display.
Max Inate:
Max Inate (repositioning mode) will rapidly inate the mattress
to provide a stable surface for nursing procedures or patient
repositioning.
As a safety feature, Max Inate will automatically return back to
the previous therapy setting after 20 minutes to avoid the patient
being on peak pressures for long periods.
An orange LED will be visible on this button when it is active.
Comfort Settings:
Press the +/- button to adjust the comfort setting from -2 (softest)
to +2 (rmest).
The control unit will automatically default to 0 (medium).
All comfort settings are within a therapeutic window and adjusting
the setting will not alter clinical therapy.
The chosen comfort setting will be visible on the digital display.
Pressure Monitoring:
This LED detects changes of pressure inside the mattress. If blue,
the pressures are adjusting (e.g. inating). If green, the pressures
are static.
Note: When this LED is green on initial ination, it can be a useful
indication that the mattress is fully inated and ready for patient use.
Question and Answer IconNeed help?

Do you have a question about the Medstrom Aria PRO and is the answer not in the manual?

Summary

Control Unit Functions

Power

Ensures the rocker switch is ON, press POWER/MUTE/UNLOCK to turn on. A green LED indicates the unit is in use.

Unlock Panel

The unit locks after 3 minutes of inactivity. Press and hold POWER/MUTE/UNLOCK for 3 seconds to unlock.

Constant Immersion Mode

Default mode. Maintains constant air cell pressure for maximum contact and pressure redistribution. Green LED indicates active.

Gradient Immersion Mode

Active therapy alternating air cells for off-loading. Selectable cycle times (5, 10, 15 mins) visible on display.

Max Inflate Mode

Rapidly inflates mattress for stable surface during procedures. Auto-resets after 20 mins. Orange LED indicates active.

Comfort Settings Adjustment

Adjust comfort from -2 (softest) to +2 (firmest). Default is 0 (medium). Settings are within a therapeutic window.

Pressure Monitoring Indicator

LED indicates pressure changes. Blue means adjusting, green means static. Green on initial inflation means ready.

Alarms and Indicators

System Alarms

Alerts for power failure or abnormal pressure. Silenced by POWER/MUTE/UNLOCK. May re-alert after 3-25 minutes.

Immersion Scale

LED shows internal pressures: blue (low), green (medium), orange (high). Indicates immersion achieved based on patient.

CPR Function

CPR valve at foot-end. Twist to OPEN for rapid deflation. Twist back to CLOSE to deactivate.

Transport Mode

Self-sealing technology keeps mattress inflated up to 15 hours during power failure. Unit alarms periodically.

Bridging Cells for Heel Off-Loading

Air cells can be replaced with bridging cells (5" smaller) for off-loading critical areas like heels. Disconnect and reconnect cells.