Ensures the rocker switch is ON, press POWER/MUTE/UNLOCK to turn on. A green LED indicates the unit is in use.
The unit locks after 3 minutes of inactivity. Press and hold POWER/MUTE/UNLOCK for 3 seconds to unlock.
Default mode. Maintains constant air cell pressure for maximum contact and pressure redistribution. Green LED indicates active.
Active therapy alternating air cells for off-loading. Selectable cycle times (5, 10, 15 mins) visible on display.
Rapidly inflates mattress for stable surface during procedures. Auto-resets after 20 mins. Orange LED indicates active.
Adjust comfort from -2 (softest) to +2 (firmest). Default is 0 (medium). Settings are within a therapeutic window.
LED indicates pressure changes. Blue means adjusting, green means static. Green on initial inflation means ready.
Alerts for power failure or abnormal pressure. Silenced by POWER/MUTE/UNLOCK. May re-alert after 3-25 minutes.
LED shows internal pressures: blue (low), green (medium), orange (high). Indicates immersion achieved based on patient.
CPR valve at foot-end. Twist to OPEN for rapid deflation. Twist back to CLOSE to deactivate.
Self-sealing technology keeps mattress inflated up to 15 hours during power failure. Unit alarms periodically.
Air cells can be replaced with bridging cells (5" smaller) for off-loading critical areas like heels. Disconnect and reconnect cells.