CONTOUR® NEXT LINK Accuracy
Industry standards have an acceptable error range of ±20% to lab results.
1
Bayer’s CONTOUR® NEXT LINK meter gives even more accurate readings,
±10% to lab results.
2
1. International Organization for Standardization. ISO 15197:2003 section 7.4 requires results within ±20% range, specifically: ≥ 95% of results to fall within ±0.83 mmol/L of a laboratory reference value for
blood glucose concentrations <4.2 mmol/L and within ±20% of a laboratory reference value for blood glucose concentrations ≥4.2 mmol/L.
2. Exceeds the minimum accuracy requirement set out in the ISO 15197:2003 standard with ±10% accuracy vs. laboratory method for blood glucose concentrations ≥ 4.2mmol/L and ±0.56 mmol/L accuracy vs.
laboratory method for blood glucose concentrations < 4.2 mmol/L
Meter Readings
Lab Result Results ±20% to lab Results ±10% to lab
4.3 mmol/L 3.4 - 5.2 mmol/L 3.9 - 4.7 mmol/L
10 mmol/L 8 – 12 mmol/L 9-11 mmol/L
Examples of possible meter readings: ±20% vs. ±10% accuracy