EasyManuals Logo

Megger DET4 Series User Manual

Megger DET4 Series
63 pages
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Page #27 background imageLoading...
Page #27 background image
27
electrode under test is calculated as R=V/I2 or R=V/(I-I1). The current transducer (ICLAMP) measures I2
and feeds this value back to the instrument.
Principle of operation (two-clamp stakeless resistance measurement)
In this example, the electrode under test is connected to a network of other electrodes. It is either
impractical or unsafe to disconnect an individual electrode for testing. Also, there might be insufficient
space to perform a classic three-terminal resistance measurement. The stakeless test method using both
VCLAMP and ICLAMP can be used to obtain a measurement for the electrode under test.
A defined test voltage is injected into the system using the VCLAMP, inducing a current, I, to flow and be
measured by the ICLAMP. The model shown in Figure 7 can be simplified to the resistance of the
electrode under test, Rx and the resistance of the other electrodes in parallel, i.e. R1 || R2 || … || Rn.
Therefore, the current induced by the test voltage is I=V/[R
x
+(R
1
|| R
2
|| … || R
n
)]. It follows that as the
resistance of the other electrodes in parallel approaches zero, then the resistance measured, approaches
the value of the electrode under test.
Figure 7: schematic for two-clamp stakeless resistance measurement

Other manuals for Megger DET4 Series

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Megger DET4 Series and is the answer not in the manual?

Megger DET4 Series Specifications

General IconGeneral
BrandMegger
ModelDET4 Series
CategoryTest Equipment
LanguageEnglish

Related product manuals