EasyManuals Logo

Megger DET4 Series User Manual

Megger DET4 Series
63 pages
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Page #62 background imageLoading...
Page #62 background image
62
ICLAMP SPECIFICATIONS
Measurement range: 1 mA to 1200 A
Transformation ratio: 1000 : 1
Output signal: 1 mA / A (1 A @ 1000 A)
Accuracy
1mA to 100 mA : 3% of reading
0.1 A to 1 A : 2% of reading
1 A to 10 A : 1% of reading
10 A to 100 A : 0.5% of reading
Operating temperature
-10°C to +50°C / 14°F to 122°F
0% to 85% RH at +35°C / 95°F
Storage temperature
-20°C to +70°C / -4°F to 158°F
Jaw opening: 57 mm maximum (2.3 in)
Maximum conductor size: 52 mm (2.1 in)
Insulation: Double insulation
Voltage rating: 600 V CAT III
Current rating: 1200 A for 20 minutes
Ingress protection: IP40
Safety
EN61010-2-032
EMC
EN 50081-1 Class B
EN 50082-2
Dimensions:
111 mm x 216 mm x 45 mm (4.4 in x 8.5
in x 1.8 in)
Weight: 0.5 kg approximately (1.2lb)
VCLAMP SPECIFICATIONS
Insulation: Double insulation
Voltage rating: 600V CAT III
Current rating: 1200A for 20 minutes
Ingress protection: IP40
Operating temperature
-10°C to +50°C / 14°F to 122°F
0% to 85% RH at +35°C / 95°F
Storage temperature
-20°C to +70°C / -4°F to 158°F
Jaw opening: 57mm maximum (2.3in)
Maximum conductor size: 52mm / 2.1in
Electrical safety
EN61010-2-032
EMC
EN 50081-1 Class B
EN 50082-2
Dimensions: 111mm x 216mm x 45mm
(4.4in x 8.5in x 1.8in)
Weight: 0.5 kg approximately (1.2 lb)

Other manuals for Megger DET4 Series

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Megger DET4 Series and is the answer not in the manual?

Megger DET4 Series Specifications

General IconGeneral
BrandMegger
ModelDET4 Series
CategoryTest Equipment
LanguageEnglish

Related product manuals