Care and maintenance | EN
21
6
Place the upper section on the new collection container. Hold the collection container and turn
the upper part clockwise as far as it will go; carry out a visual inspection (all retaining lugs of the
container must be properly engaged).
new old
7 Grease the sealing surfaces in the retaining element with Vaseline.
8
Carefully push the amalgam separator back into the retaining element and close the locking clip.
9
Switch on the main switch of the treatment unit.
Theunitstartsupbriey1to2timesandLED1ontheexternaldisplaylightsupgreen(„ready
for operation“).
RESET
RESET
RESET
RESET
RESET
RESET
LEUCHTEN
BLINKEN
1
RESET
RESET
2
RESET
RESET
3
RESET
RESET
4
RESET
RESET
1
2 3
4
10
Ifthecollectioncontainerisnotinsertedcorrectly,LED2ontheexternaldisplayashesredand
a beep is heard. In this case, switch off the main switch and carefully repeat the procedure as
described above.