EN | Care and maintenance
24
1
Start service mode
Call up service mode (see 7.2.3 Service mode)
2 Perform motor and valve test run at least once
3 End service mode by switching off the supply voltage
7.2.5. 5-year service kit
i
According to the German Waste Water Ordinance, Annex 50, amalgam separators must be inspected for proper condition at intervals
of no longer than 5 years in accordance with state law!
The 5-year inspection must be entered in the device document!
1
Carry out a 1-year inspection see 7.2.4. 1-year service kit
2
Check the correct installation and connection of the amalgam separator in accordance with the
installation requirements
see 5.2. Installation requirements
3 Carry out disinfection and cleaning of the suction hoses and spittoon
see 7.1.1. Daily cleaning with GREEN&CLEAN
M2
4 Carry out a visual inspection of the centrifuge chamber see 7.2.5.1. Visual check of the centrifuge