Questions and Answers

  • M
    Mikayla DavisAug 19, 2025
    Why are my METER GROUP AquaLab 4TEV DUO Measuring Instruments readings slow or inconsistent?
    • A
      Ashley BestAug 19, 2025
      Slow or inconsistent readings on your METER GROUP Measuring Instruments could be due to several factors. The sample chamber might be dirty and require cleaning. Also, a significant temperature difference between the sample and the block chamber can cause this issue; allow the sample to reach the instrument's temperature. Some products naturally absorb or desorb moisture slowly, extending measurement times. If the sample contains volatiles, this can also affect readings. Lastly, a broken or bent fan blade in the block chamber may be the cause. If salt standards take a long time to read and the sample chamber is clean, contact Customer Support for details on replacement.
  • J
    Jennifer OchoaAug 21, 2025
    What to do if verification is not correct on METER GROUP AquaLab 4TEV DUO Measuring Instruments?
    • T
      taylordavidAug 21, 2025
      If the verification on your METER GROUP Measuring Instruments is incorrect, start by cleaning the sample chamber and mirror. If the verification is still not correct, verify and adjust for linear offset.
  • C
    Craig KruegerAug 23, 2025
    What to do if AQUALAB Measuring Instruments displays 'Remove sample. Sample is too hot!' error message?
    • C
      christine46Aug 23, 2025
      The 'Remove sample. Sample is too hot!' error message indicates that the sample temperature is too high for the instrument to equilibrate quickly. Make sure your samples are at the same temperature as the instrument.
  • M
    Michelle BruceAug 26, 2025
    How to fix high or low water activity readings on METER GROUP AquaLab 4TEV DUO verification standards?
    • C
      Crystal SmithAug 27, 2025
      If water activity readings on your METER GROUP Measuring Instruments verification standards are too high or too low, and a linear offset adjustment doesn't solve the problem, the thermopile or the chamber mirror may be contaminated and require cleaning.
  • J
    Jose GilbertAug 27, 2025
    What to do if moisture content displayed is not correct on METER GROUP AQUALAB DUO?
    • Q
      qhebertAug 28, 2025
      If the moisture content displayed on your METER GROUP AQUALAB DUO is incorrect, the selected model may not be appropriate for the product you're testing. Try toggling through the available models to find a better match. If the model is correct but still inaccurate, you might need to offset the model. If the model is correct and an offset does not work, it may be necessary to generate a new model for the product or update an existing model. For information about updating or generating a model, contact Customer Support.
  • J
    Jessica RobertsonAug 30, 2025
    What to do if sensors are reading outside of the typical range on METER GROUP Measuring Instruments?
    • A
      Amy WhiteAug 30, 2025
      If the sensors on your METER GROUP Measuring Instruments are reading outside the typical range, and cleaning and troubleshooting don't resolve the issue, go to the Diagnostics screen. The lid, base, and sample temperatures should be between 24.5 and 25.5 °C. If the mirror temperature matches the lid temperature, the cooler has likely failed and needs replacement. If the mirror temperature is below the lid temperature or seems random, the thermocouple wire is probably broken and needs repair. For optical range, a typical optical range is between 300 and 2,900 mV. If the sensor is reading outside of this range, contact Customer Support.
  • M
    Malik FletcherSep 1, 2025
    What to do if moisture content does not show up on the Measurement tab of METER GROUP AQUALAB DUO?
    • D
      Dylan SosaSep 1, 2025
      If the moisture content isn't showing on the Measurement tab of your METER GROUP AQUALAB DUO, it might not be activated. Go to the Configuration tab, select %Moisture, and choose the appropriate model. If no models are listed in the %Moisture screen, reload them using AquaLink 4. If %Moisture isn't an active selection, the DUO feature may not be enabled. Contact Customer Support to learn how to upgrade to a DUO.
  • M
    markfernandezSep 3, 2025
    How to fix 'Displayed a is below w instrument detection limits' on METER GROUP Measuring Instruments?
    • C
      Casey FrenchSep 3, 2025
      The message 'Displayed a is below w instrument detection limits' on your METER GROUP Measuring Instruments means the sample is too dry for an accurate reading. This occurs when there isn't enough moisture to condense on the mirror. Try cleaning the mirror and chamber, then measure the sample again.
  • R
    Randy DaughertySep 6, 2025
    What causes METER GROUP AQUALAB not to turn on and how to fix it?
    • V
      Victoria PachecoSep 6, 2025
      If your METER GROUP AQUALAB doesn't turn on, a power surge may have blown a fuse. First, check that the power cord is securely connected to both the instrument and the power outlet. To replace the fuse: 1. Unplug the power cord. 2. Locate the fuse box on the right side of the panel where the power cord plugs in; press the release tab and pull out the fuse-holder. 3. Replace the broken fuse with a 1.25-A, 250-V fuse. Important: Use only this type of fuse to avoid damage or voiding the warranty. 4. Replace the fuse holder and push it back in until the release tab clicks. 5. Reconnect the power cord and try turning on the instrument. If the fuse blows again, a component may have failed, and you should contact Customer Support for repairs.
  • B
    Benjamin JenkinsSep 8, 2025
    What does 'Ran test with wrong model' mean for METER GROUP AquaLab 4TEV DUO?
    • M
      mollykaiserSep 8, 2025
      If your METER GROUP Measuring Instruments displays 'Ran test with wrong model', on the measurement screen, toggle to the correct model using the UP and DOWN button. The moisture content value updates to correspond with the model selected. Use SAVE to save the updated moisture value. If the correct model is not available, the model may not be loaded on the instrument. To determine which models are loaded on the instrument, cycle to the Configuration tab, select Moisture and the loaded models will be listed. If the correct model is not available, load the appropriate model using AquaLink 4 software. The AQUALAB DUO can hold a total of 100 models at any one time.

Summary

2. Operation

2.1 Installation

Provides steps for setting up the AQUALAB instrument.

Installation Preparation

Details the preparation steps for AQUALAB installation.

2.2 Sample Preparation

2.2.1 Sample Composition

Discusses sample composition and its effect on reading times.

2.2.2 Sample Temperature

2.3 Taking a Reading

2.4 Viewing Data

2.5 Using Moisture Content (AQUALAB DUO Only)

2.5.2 Loading and Organizing Product Isotherm Models

Guides on loading and managing isotherm models via software.

2.5.3 Measuring Moisture Content

2.5.4 Moisture Content Model Offset

2.5.5 Restore Moisture Content Default Settings

3.1 Specifications

3.2 Components

3.2.4.1 Measurement Tab

3.2.4.2 Configuration Tab

Configuration Tab Options

Calibration

Select to start the calibration process.

Mode

Choose measurement mode: Single, Continuous, Custom, Low-Emitting.

Diagnostics

View sensor information and status.

3.2.4.3 Data Tab

3.3 Computer Interface

3.3.1 AquaLink 4 Software

3.3.2 Using a Terminal Program

3.4.1 Moisture Content

3.4.2 Water Activity

3.4.2.1 Temperature Effects

4. Service

4.1 Verification and Calibration

4.1.1 Water Activity Verification

Procedures for verifying water activity calibration.

4.1.1.1 Verification Standards

Lists unsaturated salt solutions used for verification.

4.1.2 Linear Offset

4.1.3 Multipoint Calibration (TEV Only)

4.1.4 Restore Factory Defaults

4.2 Cleaning

4.2.3 Cleaning Procedure

Step-by-step instructions for cleaning the instrument.

4.4 Repair

4.5 Troubleshooting

4.6 Customer Support

METER GROUP AquaLab 4TEV DUO Specifications

General IconGeneral
ModelAquaLab 4TEV DUO
Measurement TypeWater Activity and Moisture Content
Water Activity Accuracy±0.003 aw
Water Activity Resolution0.001 aw
Temperature Accuracy±0.2 °C
DisplayLCD
ConnectivityUSB
Water Activity Range0.00 to 1.00 aw
Moisture Content Range0 to 100%
Moisture Content Accuracy±0.5%
Temperature Range15°C to 50°C
Measurement TimeLess than 5 minutes
Operating Temperature15°C to 50°C
Power SupplyAC adapter
Power Requirements100-240V, 50/60Hz