29Wattmeters
L1
W
L1
L2
L2
L3 L3
U U+
I
I+
Alimentation
Charge
Fig 4.4.2 Three-phase/balanced wires (T3FE)
connection diagram (PX 120 only)
Common characteristics of power measurements
The active and reactive power measurements are signed, whereas the apparent power
measurement is not.
The display indicates “OL if the voltage or current is in “OL”.
3.4.1 Specific reference conditions
DC: AC component < 0.1% of the DC signal
AC: - Sinusoidal signals
- PF = 1
- Frequency: 50 Hz
3.4.2 Active power measurements
1. Press the ON/OFF key: the first screen appears
2. Connect the measurement leads to the instrument’s current measurement terminals
(on the left), complying with the polarities indicated: red lead on the + terminal and
black lead on the COM terminal.
Connect the measurement leads to the instrument’s voltage measurement terminals
(on the right), complying with the polarities indicated: red lead on the + terminal
and black lead on the COM terminal.
Set up as indicated in the connection diagrams in Fig.4.4.1 (single phase) or Fig.4.4.2
(balanced three-phase, T3FE only PX 120), depending on the case, making sure if
possible that the maximum acceptable limits are not exceeded (see table below).
Range switching is automatic.
Single-phase characteristics
Display range 1000 W 6 kW
Measuring range 10.00 W...999.9 W 1000 W...5999 kW
Accuracy at 50Hz for PF 0.8 1.5% of reading ±2 cts 1.5% of reading ±2 cts
Accuracy with DC 2.5%of reading ±5 cts 2.5%of reading ±5cts
Resolution 0.1W 1 W
Three-phase characteristics
Display range 1000 W 6 kW
Specified measuring range 10.00 W - 999,9 W 1000 W – 5999 W
Accuracy at 50Hz for PF 0.8 3% of reading ±2 cts 3% of reading ±2 cts
Resolution 0.1W 1 W
Note: With three-phase, the measurement of the active power is only performed
on the fundamental components, so it is only valid for sinusoidal signals.
Chapter III
Power supply
Load
北京海洋兴业科技股份有限公司(证券代码:839145)
电话:010-62176775

Other manuals for Metrix PX 120

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Metrix PX 120 and is the answer not in the manual?

Metrix PX 120 Specifications

General IconGeneral
BrandMetrix
ModelPX 120
CategoryTest Equipment
LanguageEnglish

Related product manuals