33Wattmeters
4.3 Environmental parameters
■■
■■
■ Climatic conditions
Temperature in °C
95
90
75
45
10
-40
0
20
26
35
50
70
➂
➂
➁
➀
Relative humidity in % RH
Note: The operating range (2) is reduced to 40°C with the mains adapter accessory.
■■
■■
■ Altitude
- Operation: ≤ 2,000 m
- Storage ≤ 10,000 m
■■
■■
■ Impermeability:protection index IP 54 (according to EN 60529)
(IP2X electrical protection for the terminals)
4.4 Compliance with norms
■■
■■
■ Electrical safety
(as per EN 61010-1)
- Dual insulation:
- Installation category: III
- Pollution level: 2
- Rated voltage: 600 V
■■
■■
■ Electromagnetic compatibility (EN 61326-1)
■■
■■
■ Electrostatic discharge (IEC 61000-4-2)
- 4 kV contact, 8 kV air
- aptitude criterion A, industrial site
■■
■■
■ Electromagnetic field (IEC 61000-4-3)
- 10 V/m
- aptitude criterion A, industrial site
■■
■■
■ Bursts (IEC 61000-4-4)
- 2 kV Level III
- aptitude criterion B, industrial site
■■
■■
■ Shock wave (IEC 61000-4-5)
- 1 kV differential mode, 2 kV common mode
- aptitude criterion A, industrial site
■■
■■
■ Conducted interference (IEC 61000-4-6)
-3 V
- aptitude criterion A, industrial site
■■
■■
■ Mechanical resistance
- Operating position: any - Impacts: 100 g (test as per IEC 68-2-27)
- Free fall: 1 m (test as per IEC 68-2-32) - Vibrations: 0.15 mm (test as per IEC 68-2-6)
■■
■■
■ Auto-extinguishability
- Housing: V0
- Display window: V0
Chapter IV
➀ Reference range
➁ Operating range
➂ Storage range (without battery)
北京海洋兴业科技股份有限公司(证券代码:839145)