EasyManua.ls Logo

Metrix ST5484E - Page 9

Metrix ST5484E
12 pages
Print Icon
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
Doc# M9162• REV AB (April 2019) Page 9 of 12
7. INSTALLATION BEST PRACTICES FOR RFI IMMUNITY
The ST5484E is case isolated from the loop power connecon to the
PLC, DCS, SCADA, or other +24 VDC alarm system. With the correct
grounding scheme, this type of isolaon avoids ground loops and re-
sults in good RF immunity. However, in situaons which involve inter-
ference from high-powered transmiers and especially spread-spec-
trum type radios, the ST5484E may become vulnerable to RF noise.
Since it is oen dicult to idenfy these situaons in advance, Metrix
strongly recommends ulizing the following Best Pracces when install-
ing all ST5484E transmiers in the eld:
1. Twist the sensor leads together (at least 10 twists per foot).
2. Slide the twisted leads through the included ferrite core and posi-
on as close to the sensor as praccal. Refer to Figures 4 and 5.
3. If possible, cut the leads short and use a shielded twisted pair. Fol-
low step 2 in this applicaon as well.
As shown in Figure 5, chokes” that aenuate dierent frequencies can
be created by looping a dierent number of turns through the ferrite
core. More wire turns adds more inductance, aenuang progressively
lower frequencies. Mulple “chokes” can be created in this manner
and placed in series on the wire leads to cover a broad frequency band.
Always locate the choke covering the higher frequency band closest to
the sensor.
8. CALIBRATION
The ST5484E transmier has been factory calibrated for the full-scale vibraon level marked
on the label. If the calibraon is in doubt, the unit can be veried in the eld by following
the procedures outlined below. Note that there are no Zero and Span adjustments on the
transmier. Addionally, the transmier uses a true RMS amplitude detecon circuit; units
supplied with a full scale range in peak units scale the underlying RMS measurement by a
Figure 5: Leads may also be
looped through the ferrite
core to aenuate dierent
frequencies. The opmal
number of loops will vary
by applicaon and the
interfering frequency(ies).
Figure 4: Leads may be
simply fed straight through
the ferrite core.
factor of 1.414 to provide a “derived peak” rather than true peak measurement.
7.1 Zero Vericaon
In the absence of vibraon the output current should be 4 mA ± 0.1 mA. If the ambient
vibraon exceeds 2% of full scale, the transmier should be removed from the machine and
placed on a vibraon free surface for this measurement. Oen a piece of foam can be used
to isolate the transmier from external moon.
7.2 Span Vericaon
Subject the transmier to a known vibraon within the full scale range marked on the label.
If you are using a portable vibraon shaker where it can be tested at full scale, the output
should be 20 mA ± 0.5 mA.
9. OPTIONAL DYNAMIC OUTPUT
The transmier can be ordered with an oponal dynamic output. Transmiers supplied with
this opon can be idened by the presence of four ying leads instead of two (Figure 6), or
four terminal connecons instead of two (Figure 7).
The dynamic output is an acceleraon signal with a sensivity of 100 mV/g, ltered to the

Other manuals for Metrix ST5484E