Technical specifications
■■■■■■■■
19
Instrument status min. 30 minutes in
operation
Adjusting interval annual
Measuring accuracy applies for all measur-
ing ranges without sen-
sor error, under refer-
ence conditions, meas-
uring interval 100 ms
3.18 OMNIS Titrator – Liquid Handling specifications
Dosing drive
Dosing resolution 100,000 steps per cylinder vol-
ume
Dosing accuracy 0.01% typically, of the dosed
volume with respect to
the cylinder volume
3.19 Magnetic stirrer – Specifications
OMNIS product version
with attached magnetic
stirrer
Adjustment range for rotational
speed
+1 to +15 Rotation in counter-
clockwise direction
(seen from above)
–1 to –15 Rotation in clockwise
direction (seen from
above)
Rotational speed change per step 120 rpm
Maximum rotational speed 1,800 rpm
Stirring bar lengths 8, 12, 16, 25, 30 mm