SA 35 and SA 36
Operation and Maintenance Manual
Manual 118 Issue 15
Page 12 of 28
-2C
There are no restrictions for land, air, or sea transport.
3.13 Storage
SA suction units must be stored in a dry, dust-free, well ventilated environment. The storage
environment should not exceed the temperature and humidity conditions stated below. Avoid direct
sun or UV exposure and shield nearby sources of heat. The equipment should be stored in its original
packaging providing no damage is evident. Protect against ground moisture by storing on a shelf or
wooden pallet.
Temperature max +60°C min -20°C Humidity max +80% min +30%
-2C
3.14 Long Term Storage
When the units are held in storage, or used very infrequently, the three-monthly maintenance period
may be extended to twelve months. Particular care must be taken when inspecting flexible
components, such as valves, diaphragms, to ensure embrittlement has not occurred.
SA suction units should be stored in a cool, dry environment, as stated above.
3.15 Instructions by Medical Staff to Patients
When the equipment is required by a patient for home use, Medical Staff must fully instruct the patient
on the safe operation of the equipment. In the event of equipment contamination or failure, the patient
must be advised to switch off the unit and contact the authority from which the unit is loaned.
3.16 Troubleshooting
Problem: Cause: Solution:
No power to unit
Unit not turned on Turn Main Switch „ON/OFF” ON
Wrong operating voltage Check mains voltage output
Mains not connected Connect mains cable
Defective Fuse
Check fuse and replace if
indicated
Pumps fail to run
Leak in vacuum system Check all connection and tubing.
Collection Jar is full Replace Jar
Pump suction power too weak
Vacuum leaks
Check all seals and hoses. Make
sure Lid is securely on Jar.
Vacuum not to required setting
Turn vacuum control clockwise
until the desired suction power is
reached
Tubing is plugged, bent and/or
disconnected
Replace tubing if plugged and
eliminate any bends
Internal hydrophobic filter is
blocked or wet
Replace internal hydrophobic
filter. – Must be done by EBME
External filter is blocked Replace External filter.
If a problem cannot be solved, contact a fully trained and qualified engineer (EBME), authorized MGE
dealer or MGE (sales@mgelectric.co.uk).