1. SAFETY INSTRUCTIONS
The safety of the patient and suction unit operator are the first priorities. It is therefore vital that the
following precautions are strictly observed:
WARNING!!!
Never operate a SAM suction unit without an MGE filter capsule.
The filter capsules are single patient use and cannot be reused. Reuse of such devices can be
dangerous for both patient and operator.
A hydrophobic filter must be fitted internally to prevent liquid passing through to the pump.
Hydrophobic filters only work once and should be replaced immediately it becomes wetted.
No modification of this equipment is allowed.
Only original and approved spare parts and collection systems must be used with all SAM suction
units – failure to use original or approved spares will invalidate the warranty and may cause injury or
damage the unit.
No liability can be accepted by MGE for units affected by the occurrence of overflow when Disposable
Liner Systems are being used.
Other than for routine daily procedures, any maintenance or repairs to MGE products must be carried
out by fully trained and qualified Electro-Biomedical engineer/technician (EBME) or an authorized
MGE dealer. Such persons must be familiar with the relevant standards, rules, accident prevention
regulations, and operating conditions as a result of their training, experience, and instruction. They
are qualified to carry out the required activities and in doing so recognize and avoid potential hazards.
All testing on SA suction units should be in accordance with ISO 10079-1
Contamination may be present on any components. When cleaning or replacing any part of the SA
unit appropriate protective clothing and gloves MUST be worn to avoid contamination. Disposal of
contaminated parts must be according to local protocols.
Store the manual in a safe place, so that it is available to the trained personnel at all times.
All collection containers must be securely mounted when in use.
The overflow valve may not operate fully against frothing. To prevent frothing anti-foam agent maybe
used.
When replacing a full collection container, be aware of its weight and ensure handling the container is
comfortable to avoid the possibility of spillage.
Transport of the SA suction unit with a full jar attached is not advised.
Solvent-based cleaning agents or abrasive cleaners must not be used on any SA suction units. Do
not wash any SA suction units under running water or submerge in water.
Any dismantling and re-assembly of this equipment - for whatever purpose - must be followed by
testing in accordance with the manufacturers’ recommendations as specified for monthly
maintenance.