Technical Data
thermoIMAGER TIM
2.5 Measurement Specifications
Model TIM 160 200 230 400 450 640
Temperature ranges -20 ... 100 °C; 0 ... 250 °C; 150 ... 900 °C; option 200 ... 1500 °C
1
Detector UFPA, 160 x 120 Pixel UFPA, 382 x 288 Pixel UFPA, 640 x
480 Pixel
Spectral range 7.5 ... 13 μm
Lenses (FOV) 23 ° x 17 °; 6 ° x 5 °; 48 ° x 31 °; 72 ° x 52 °
2
38 ° x 29 °; 13 ° x 10 °; 62 ° x 49 ° 33 ° x 25 °
System accuracy
3
±2 °C or ±2 %
Temperature
resolution (NETD)
0.08 K with 23 °;
0.3 K with 6 °;
0.1 K with 48 ° and 72 °
0.08 K with 38 °
and 62 °;
0.1 K with 13 °
4
0.04 K with 38 °
and 62 °; 0.06 K
with 13 °
4
0.075 K with
33 °
4
Frame rate 120 Hz 128 Hz
5
128 Hz 80 Hz
Warm-up time 10 min
Emissivity 0,100 ... 1,000
Visual camera - 640 x 480
pixels, 32 Hz,
54 ° x 40 ° FOV
640 x 480
pixels, 32 Hz,
30 ° x 23 ° FOV
- - -
Software TIM Connect
1) The additional measurement range of 200 … 1500 °C is not available for cameras thermoIMAGER TIM 450 and 640 as well as for
camera version TIM 160 / TIM 200 featuring 72 ° HFOV optic.
2) We recommend the lenses with 48 ° optic for the camera thermoIMAGER TIM 200 and the lenses with 23 ° optic for the camera
thermoIMAGER TIM 230 for ideal combination of IR and VIS image.
3) At ambient temperature 23±5 °C; whichever is greater.
4) Value is valid at 40 Hz and 25 °C room temperature
5) The following options can be set:
Option 1 (IR with 96 Hz at 160 x 120 px; VIS with 32 Hz at 640 x 480 px);
Option 2 (IR with 128 Hz at 160 x 120 px; VIS with 32 Hz at 596 x 447 px)