6
battery displayed). Although the device will continue to measure
reliably, you should obtain replacement batteries.
Flat battery - replacement
When the batteries are flat, the battery symbol AN will flash as soon
as the device is switched on (flat battery displayed). You cannot
take any further measurements and must replace the batteries.
1. Open the battery compartment 5 on the bottom of the device.
2. Replace the batteries ensure correct polarity as shown by the
symbols in the compartment.
Which batteries and which procedure?
Using rechargeable batteries
You can also operate this device using rechargeable batteries.
7. Using a mains adapter
You can operate this device using the Microlife mains adapter
(DC 6V, 600 mA).
1. Plug the adapter cable into the mains adapter socket 4 in the
blood pressure monitor.
2. Plug the adapter plug into the wall socket.
When the mains adapter is connected, no battery current is
consumed.
8. Error Messages
If an error occurs during the measurement, the measurement is
interrupted and an error message, e.g. «Err 3», is displayed.
Use 4 new, long-life 1.5 V, size AAA alkaline batteries.
Do not use batteries beyond their date of expiry.
Remove batteries if the device is not going to be used for a
prolonged period.
Only use «NiMH» type reusable batteries.
Batteries must be removed and recharged when the flat
battery symbol appears. They should not remain inside the
device as they may become damaged (total discharge as a
result of low use of the device, even when switched off).
Always remove the rechargeable batteries if you do not
intend to use the device for a week or more.
Batteries cannot be charged in the blood pressure monitor.
Recharge batteries in an external charger and observe the
information regarding charging, care and durability.
Only use the Microlife mains adapter available as an
original accessory appropriate for your supply voltage.
Ensure that neither the mains adapter nor the cable are
damaged.
Error Description Potential cause and remedy
«Err 1» Signal too
weak
The pulse signals on the cuff are too
weak. Re-position the cuff and repeat
the measurement.*
«Err 2»
BK
Error signal During the measurement, error signals
were detected by the cuff, caused for
instance by movement or muscle
tension. Repeat the measurement,
keeping your arm still.
«Err 3»
BT
No pressure
in the cuff
An adequate pressure cannot be
generated in the cuff. A leak may have
occurred. Check that the cuff is
correctly connected and is not too
loose. Replace the batteries if neces-
sary. Repeat the measurement.
If the 2-year warranty period has
expired, it is recommended to replace
with a new Microlife cuff available at
your pharmacy.
«Err 5» Abnormal
result
The measuring signals are inaccurate
and no result can therefore be
displayed. Read through the checklist
for taking a reliable measurement and
then repeat the measurement.*
«Err 6» AFIB/MAM
Mode
There were too many errors during the
measurement in AFIB/MAM mode,
making it impossible to obtain a final
result. Read through the checklist for
performing reliable measurements and
then repeat the measurement.*

Table of Contents

Other manuals for Microlife AFIB ADVANCED Easy

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Microlife AFIB ADVANCED Easy and is the answer not in the manual?

Microlife AFIB ADVANCED Easy Specifications

General IconGeneral
BrandMicrolife
ModelAFIB ADVANCED Easy
CategoryBlood Pressure Monitor
LanguageEnglish

Related product manuals