7
Using rechargeable batteries
You can also operate this instrument using rechargeable batteries.
9. Error Messages
If an error occurs during the measurement, the measurement is
interrupted and an error message, e.g. «ERR 3», is displayed.
* Please consult your doctor, if this or any other problem occurs
repeatedly.
10. Safety, Care, Accuracy Test and Disposal
Safety and protection
)
Only use «NiMH» type reusable batteries.
)
Batteries must be removed and recharged when the flat
battery symbol appears. They should not remain inside the
instrument as they may become damaged (total discharged as
a result of low use of the instrument, even when switched off).
)
Always remove the rechargeable batteries if you do not
intend to use the instrument for a week or more.
)
Batteries cannot be charged in the blood pressure monitor.
Recharge batteries in an external charger and observe the
information regarding charging, care and durability.
Error Description Potential cause and remedy
«
ERR 1
» Signal too
weak
The pulse signals on the cuff are too
weak. Re-position the cuff and repeat the
measurement.*
«
ERR 2
» Error signal During the measurement, error signals
were detected by the cuff, caused for
instance by movement or muscle tension.
Repeat the measurement, keeping your
arm still.
«
ERR 3
» No pressure
in the cuff
An adequate pressure cannot be gener-
ated in the cuff. A leak may have
occurred. Check that the cuff is correctly
connected and is not too loose. Replace
the batteries if necessary. Repeat the
measurement.
«
ERR 5
» Abnormal
result
The measuring signals are inaccurate and
no result can therefore be displayed.
Read through the checklist for performing
reliable measurements and then repeat
the measurement.*
«
HI
» Pulse or cuff
pressure too
high
The pressure in the cuff is too high (over
300 mmHg) OR the pulse is too high (over
200 beats per minute). Relax for 5
minutes and repeat the measurement.*
«
LO
» Pulse too low The pulse is too low (less than 40 beats
per minute). Repeat the measurement.*
)
If you think the results are unusual, please read through the
information in «Section 1.» carefully.
This instrument may be used only for the purpose described in
this booklet. The manufacturer cannot be held liable for damage
caused by incorrect application.
This instrument comprises sensitive components and must be
treated with caution. Observe the storage and operating condi-
tions described in the «Technical Specifications» section!
Protect it from:
water and moisture
extreme temperatures
impact and dropping
contamination and dust
direct sunlight
heat and cold
The cuffs are sensitive and must be handled with care.
Only pump up the cuff once fitted.
Do not use the instrument close to strong electromagnetic fields
such as mobile telephones or radio installations.
Do not use the instrument if you think it is damaged or notice
anything unusual.
Never open the instrument.
If the instrument is not going to be used for a prolonged period
the batteries should be removed.
Read the additional safety instructions in the individual sections
of this booklet.
Ensure that children do not use the instrument unsuper-
vised; some parts are small enough to be swallowed.
Error Description Potential cause and remedy

Table of Contents

Other manuals for Microlife BP A130

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Microlife BP A130 and is the answer not in the manual?

Microlife BP A130 Specifications

General IconGeneral
BrandMicrolife
ModelBP A130
CategoryBlood Pressure Monitor
LanguageEnglish

Related product manuals