mode or 0.1 °C / 32.2 °F in object mode.
Ambient temperature too high BK: Displays «H» and when the ambient temperature is higher than 40.0 °C / 104.0 °F.
Ambient temperature too low BL: Displays «L» and when the ambient temperature is lower than 15.0 °C / 59.0 °F in body mode or
lower than 5.0 °C / 41.0 °F in object mode.
Error function display BM: The system has a malfunction.
Blank display BN: Check if the batteries have been inserted correctly. Also, check polarity (<+> and <->) of the batteries.
Flat battery indicator BO: If only icon is shown on the display, the batteries should be replaced immediately.
12. Cleaning and Disinfecting
Use an alcohol swab or cotton tissue moistened with alcohol (70% Isopropyl) to clean the thermometer casing and the measuring sensor.
Ensure that no liquid enters the interior of the device. Never use abrasive cleaning agents, thinners or benzene for cleaning, and never
immerse the device in water or other cleaning liquids. Take care not to scratch the surface of the sensor lens and the display.
13. Battery Replacement
This device is supplied with 2 new, long-life 1.5V, size AAA batteries. Batteries need replacing when this icon BO is the
only symbol shown on the display. Remove the battery cover BR by sliding it in the direction shown. Replace the batteries – ensure
correct polarity as shown by the symbols in the compartment.
Batteries and electronic devices must be disposed of in accordance with the locally applicable regulations, not with domestic waste.
14. Guarantee
This device is covered by a 5 year guarantee from the date of purchase. During this guarantee period, at our discretion, Microlife will
repair or replace the defective product free of charge. Opening or altering the device invalidates the guarantee.
The following items are excluded from the guarantee:
• Transport costs and risks of transport.
• Damage caused by incorrect application or non-compliance with the instructions for use.
• Damage caused by leaking batteries.
• Damage caused by accident or misuse.
• Packaging/storage material and instructions for use.
• Regular checks and maintenance (calibration).
• Accessories and wearing parts: Batteries, Probe cover (optional).
Should guarantee service be required, please contact the dealer from where the product was purchased or your local Microlife service.
You may contact your local Microlife service through our website: www.microlife.com/support
Compensation is limited to the value of the product. The guarantee will be granted if the complete product is returned with the original
invoice. Repair or replacement within the guarantee does not prolong or renew the guarantee period. The legal claims and rights of
consumers are not limited by this guarantee.
15. Technical Specifications
Type: Non Contact Thermometer NC 150 BT
Measurement range:
Body mode: 34.0 – 43.0 °C / 93.2 – 109.4 °F
Object mode: 0.1 – 99.9 °C / 32.2 – 211.8 °F
Resolution: 0.1 °C / °F
Measurement
± 0.2 °C, 35.0˜42.0 °C / ˜0.4 °F, 95.0˜107.6 °F
± 0.3 °C, 34.0 ˜ 34.9 °C and 42.1˜43.0 °C /
± 0.5 °F, 93.2 ~ 94.8 °F and 107.8 ~ 109.4 °F
Object mode:
±1.0 °C, 0.1 ~ 99.9 °C / ±2 °F, 32.2 ~ 211.8 °F
Display: Liquid Crystal Display, 4 digits plus special icons
Acoustic:
The unit is turned ON and ready for the measurement: 1 short beep. Complete the
measurement: 1 long beep (1 sec.) if the reading is less than 37.5 °C / 99.5 °F, 10 short
«beep» sounds if the reading is equal to or greater than 37.5 °C / 99.5 °F.
System error or malfunction: 3 short «bi» sounds.
Memory: 30 readings recall in the memory mode with a record of both time and date.