EasyManua.ls Logo

Miele professional PG 8535 - Maintenance; Routine Checks

Miele professional PG 8535
68 pages
Print Icon
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
Periodical maintenance for this lab
washer is offered by Miele Professional
Technical Service.
After 1000 operating hours or at least
once a year is the recommended
interval.
The maintenance covers the following:
–
Electrical safety
–
Door mechanism and door seal
–
Threads and connections in the
wash cabinet
–
Water inlet and drain
– Internal and external dispensing
systems
– Spray arms
– Triple Filter System
– Drain pump and non-return faucet
– Steam condenser:
– All mobile units, baskets and inserts
If available:
–
Connected printer
Part of the maintenance is a function
check:
–
A program is run as a test
–
A thermoelectrical measurement
–
A leak tightness check
–
All safety relevant measuring
systems (Display of faults)
Contact Miele Professional
Technical Service at 800-991-9380
for details.
Routine checks
The routine checks must be done daily
by the user before using the machine.
For the routine checks a checklist is
equipped with the washer.
Check the following points:
–
the filters in the base of the wash
cabinet,
–
the spray arms in the machine and at
the mobile units and inserts,
–
the wash cabinet and the door seal,
–
the mobile units, baskets and inserts,
and
– all safety equipment.
Maintenance
45