Questions and Answers

  • J
    Joseph BowmanAug 15, 2025
    How to troubleshoot a Millipore Mixer when the motor/mixing head does not rotate?
    • C
      Christopher Jacobs MDAug 15, 2025
      If the motor or mixing head of your Millipore mixer isn't rotating, first check the power supply to ensure it's properly connected. If that's not the issue, the overload protection might have tripped; in this case, check the setting amp value and the ambient temperature. Also, a heavy load of particles stuck to the magnets can prevent rotation, so clean the unit thoroughly.
  • M
    Matthew BarnesAug 18, 2025
    What to do if the outer drive head scratches the tank plate on my Millipore NovAseptic GMP 2000 Mixer?
    • W
      Wendy PerezAug 19, 2025
      If the outer drive head of your Millipore mixer is scratching the tank plate, several factors could be at play. The shaft or hole-shaft in the gearbox might be bent, in which case you should contact Millipore service. Ensure the drive unit is correctly installed. Particles stuck to the magnets can also cause scratching, so remove them. Review the tank plate welding instructions and verify tolerances if deformation occurred during welding. Lastly, check if the outer drive head is loose; if so, position it onto its lower end against the shaft and tighten it.
  • W
    William ColemanAug 21, 2025
    Why is the mixing head scratching against the tank plate on my Millipore NovAseptic GMP 2000 Mixer?
    • K
      Kevin CalderonAug 21, 2025
      If the mixing head of your Millipore mixer is scratching against the tank plate, it could be due to several reasons. The male bearing might be worn out and needs replacement. The female bearing holder could be damaged, requiring repair of the mixing head. A loose male bearing can also cause this issue; remove it, inspect for damage, and reinstall if undamaged, checking the rotation direction. Particles stuck to the magnets can also cause scratching, so remove them. Finally, deformation after welding can be a cause; review the welding instructions and verify tolerances.
  • D
    David FisherAug 25, 2025
    What causes magnetic disconnection (decoupling) in a Millipore NovAseptic GMP 2000?
    • J
      jackson29Aug 25, 2025
      If your Millipore mixer experiences magnetic disconnection of the mixing head, also known as decoupling, it could be due to several reasons. Review the guidelines under “Operation/Starting the Mixer” to ensure correct addition of substances and avoid dry running. Unexpected changes in viscosity or density can cause this; consult your Millipore representative. Ensure the rotational direction of the mixing head is correct by inspecting and reinstalling the male bearing. Increasing the ramp time can address short acceleration/retardation times. Lastly, remove any particles stuck to the magnets.
  • D
    Dr. Katherine Zimmerman DDSAug 27, 2025
    Why does my Millipore NovAseptic GMP 2000 Mixer show a high amp value?
    • S
      sarah56Aug 27, 2025
      If your Millipore mixer shows a high amp value, it could be due to a high ambient temperature; try chilling the environment. Alternatively, the motor may be mechanically overloaded; check all transmissions.
  • R
    Ryan HortonAug 30, 2025
    Why is the mixing head running eccentrically on my Millipore NovAseptic GMP 2000 Mixer?
    • C
      Caroline RodgersAug 30, 2025
      If the mixing head on your Millipore mixer is running eccentrically, it could be because the male bearing is not properly in place. Remove the male bearing, inspect it and the tank plate for damage. If undamaged, reinstall the bearing and check the rotation direction of the mixing head. Also, worn out bearings can cause this issue, so replace the male bearing and/or repair the mixing head.
  • A
    Amanda ClarkSep 2, 2025
    What causes an abnormal noise in my Millipore NovAseptic GMP 2000 Mixer?
    • C
      Cody RiceSep 2, 2025
      If your Millipore mixer is making an abnormal or rumbling noise, it might be due to a deep vortex; reduce the speed and ensure the mixing head is fully covered with liquid. Cavitation could also be the cause; investigate the pressure, temperature, and speed. Another potential cause is dry running; review the guidelines under “Operation /Starting the Mixer”.
  • A
    aliuSep 5, 2025
    Why is my Millipore NovAseptic GMP 2000 experiencing a loss of running speed?
    • C
      Charles McclainSep 5, 2025
      If your Millipore mixer is losing running speed, check the power supply to ensure correct incoming power. Also, check the fuses to ensure no phases are missing.
  • S
    Seth SmithSep 8, 2025
    What to do if the speed counter shows an incorrect number of signals on my Millipore Mixer?
    • Z
      Zachary PerezSep 8, 2025
      If the speed counter on your Millipore mixer shows an incorrect number of signals, ensure that the speed sensor unit is completely aligned and in contact with the drive unit flange without any gap. Refer to speed counter user guide 00101788PU for detailed instructions.
  • A
    Angela FernandezSep 12, 2025
    What to do if the speed counter shows no signal on my Millipore NovAseptic GMP 2000?
    • D
      Dominique WestSep 12, 2025
      If the speed counter on your Millipore mixer shows no signal, it could be due to corroded socket pins or a worn cable. In either case, replace the speed counter.

Summary

Operator and Equipment Safety

Installation and Operating Parameters

Installation

Male Bearing Installation

Step-by-step instructions for installing the male bearing.

Mixing Head Installation

Step-by-step instructions for installing the mixing head.

Drive Unit Installation

Step-by-step instructions for installing the drive unit.

Electrical Installation

Guidelines for electrical connections and setup for the mixer.

Operation

Starting the Mixer

Procedure for safely starting the NovAseptic GMP Mixer.

Operating Notes

Key considerations and best practices for operating the mixer.

Cleaning and Sterilizing Procedures

Cleaning in Place (CIP)

Procedures for performing Cleaning in Place (CIP) on the mixer.

Sterilizing in Place (SIP)

Procedures for performing Sterilizing in Place (SIP) on the mixer.

Commissioning

Troubleshooting