A - 1
A Specifications
A.1 Safety Specifications
A.1.1 Safety Classifications
The equipment is classified, according to IEC 60601-1:
A.1.2 Environmental Specifications
• The equipment may not meet the performance specifications if stored or used outside the specified
temperature and humidity ranges. If the performance of the equipment is degraded due to aging or
environmental conditions, contact your service personnel.
• When the equipment and related products have differing environmental specifications, the
effective range for the combined products is that range which is common to the specifications for all
products.
• The environmental specification of unspecified modules are the same as those of the main unit.
Type of protection against electrical shock Class I
Degree of protection against electrical
shock
Type BF defibrillation proof for CO
2
monitoring and external defibrillation.
Type CF defibrillation proof for ECG, SpO
2
, NIBP and internal defibrillation.
Degree of protection against harmful
ingress of solid
Degree of protection against harmful
ingress of water
Main unit: IP55, IP52 (with an external power supply)
Degree of safety of application in the
presence of flammable anesthetic mixture
with air or with oxygen or nitrous oxide
The equipment is not suitable for use in the presence of a flammable
anesthetic mixture with air or with oxygen or nitrous oxide
Mode of operation Continuous
Degree of mobility Portable
Main unit
Item Tempe ra tu re Relative humidity Barometric
Operating
condition
When configured
with ECG and manual
defibrillation, without
batteries
-20ºC to 55ºC 5% to 95%,
non-condensing
-382m to 4575m
(57.0 kPa to 106.2kPa)
When configured
with all functions
0ºC to 50ºC
Storage condition -40ºC to 75ºC
Shock