Status indicator
Steady green:
¡
external power supply is connected, and
the equipment operates properly.
¡
only the battery is connected, the
equipment is turned on and operates
properly.
Flashing green: only the battery is connected, the equipment
is turned off and operates properly.
Flashing red:
¡
auto test fails, or a failure is detected on the
equipment.
¡
only the battery is connected, and the
battery has a low power (battery symbol
) or battery fails (battery symbol
).
¡
only external power supply is connected,
No Battery is set to Status Indicator On
on the equipment.
Off external power supply and battery are not
connected.
Cleaning and disinfection
Make sure to clean and disinfect the equipment before
maintenance.
¡
Commonly recommended cleansers: water, ethanol (70%)
¡
Commonly recommended disinfectants: DIAN’ERKANG
surface wipes, ethanol (70%), Sodium hypochlorite bleach
(0.5%), JIAN ZHI SU surface disinfectant spray
Maintenance
A shift test should be performed at each shift (at least once a
day) or after each use. For details, consult your institution or refer
to the operator’s manual. You should turn off the equipment and
connect it to external power supply and battery after the shift
test.
A user test should be performed every week, or when the status
indicator flashes in red. To perform the user test, follow this
procedure.
1. Press the power switch to turn on the equipment.
2. Select the Main Menu quick key
→
from the Common
column select User Test.
3. If any fault exists, clear it following the displayed instructions.
4. Select User Test.
5. Perform operations following the displayed instructions, until
the equipment prompts that the user test is passed.
This guide is intended to provide a quick training only. It can not be used as
a replaement to the operator’s manual.
046-025919-00(1.0)