Do you have a question about the Mindray iMEC10 and is the answer not in the manual?
| Brand | Mindray |
|---|---|
| Model | iMEC10 |
| Category | Medical Equipment |
| Language | English |
Provides essential safety information including dangers, warnings, and cautions for safe operation and maintenance of the patient monitor.
Highlights critical warnings related to high voltage, equipment handling, and disposal procedures to prevent injury or damage.
Details precautions for avoiding electromagnetic interference and protecting the equipment from physical damage during servicing.
Introduces the patient monitor's design, capabilities, and primary functions for monitoring physiological parameters.
Explains the various system connections, including installation support and connectors for peripheral devices.
Describes the main unit's architecture, comprising front and rear housing assemblies and their key components.
Details the components and functionality of the front housing assembly, including the main board and display.
Provides instructions on how to unpack the patient monitor and check the included accessories and packing list.
Outlines essential preparation steps for installation, including site requirements and electrical considerations.
Introduces the capabilities for upgrading parameter modules, functional assemblies, and software for the iMec patient monitors.
Details hardware upgrade procedures for Wi-Fi, Recorder, CO2, IBP, and VGA/USB/Multi-functional interfaces.
Details the upgrade methods for the IBP (Invasive Blood Pressure) module.
Step-by-step guide on how to perform software upgrades on the patient monitor.
Covers procedures for upgrading various hardware components of the patient monitor, such as SD cards and touchscreens.
Provides instructions for upgrading the Wi-Fi module, including package details and upgrade methods.
Covers the upgrade process for the VGA, USB, and multi-functional interface ports.
Details the process for upgrading the 1G SD storage card for different monitor configurations.
Details the process for upgrading the recorder component, including specific procedures for different models.
Provides an overview of software upgrade procedures and the necessary tools.
Explains the procedure for upgrading the touchscreen functionality on the patient monitor.
Explains the upgrade procedure for CO2 modules, covering sidestream, mainstream, and microstream variants.
Instructions for installing the Mindray Patient Monitor Software Upgrade Tool.
Introduces the chapter on testing and maintenance procedures for ensuring the patient monitor's optimal performance.
Details performance tests for C.O. and Mainstream CO2 modules.
Describes the procedure for testing the nurse call function of the patient monitor.
Details the process of visually inspecting the equipment for any signs of damage or wear.
Provides procedures for conducting ECG performance tests and calibration.
Details NIBP tests, including leakage and accuracy checks.
Explains how to perform Cardiac Output (C.O.) tests.
Details how to test the analog output performance of the patient monitor.
Describes the procedure for performing a power-on test to verify the monitor's startup functionality.
Details the procedures for performing respiratory tests on the patient monitor.
Describes the procedure for performing temperature tests on the monitor.
Provides procedures for testing mainstream CO2 modules.
Refers to electrical safety inspection procedures for the patient monitor.
Covers advanced maintenance procedures accessible via the factory maintenance menu.
Covers performance testing procedures for various modules integrated into the patient monitor.
Explains how to perform SpO2 tests, including measurement validation.
Provides methods for IBP performance testing and calibration.
Details testing and calibration for sidestream and microstream CO2 modules.
Provides instructions for calibrating the touchscreen interface of the patient monitor.
A comprehensive guide to troubleshooting common symptoms and their possible causes and corrective actions.
Provides step-by-step instructions for disassembling the main unit of the patient monitor.
Lists parts and exploded views for the iMEC12/iMEC7 models.
Lists parts and exploded views for the iMEC15/iMEC15S models.
Lists parts and exploded views for the iMEC8/iMEC5 models.
Lists parts and exploded views for the iMEC10/iMEC6 models.
Details the test procedure and limits for mains voltage applied to parts leakage.
Details the procedure and criteria for testing protective earth resistance.
Describes how to measure patient auxiliary current and its RMS limits.
Provides procedures and limits for performing earth leakage tests on the device.
Provides a form for recording scheduled electrical safety inspection results.
Describes how to measure patient leakage current and its RMS limits.
Specifies test items for electrical safety inspection after equipment repair.