7−7. Detection of stability
The function to detect stability is judged that the measuring value is steady when
the status that the change width of total load value is within the range set by the
data width to detect stability, continues longer than the time set by the time to de-
tect stability.
7−7−1. Range to detect stability
Execute the setting of range to detect stability with the function F−65.
It can be set the range of [00000] 〜 [00099] every [0.5]D.
OFF when set the [00000]. Default is set as [00010].
Detection of stability data range of the set value per [n] is calculated the method below in case
of showing in display.
[detection of stability data range]= 0.5D × [detection of stability range coefficient n] × [scale
interval]
Setting of detection of stability range coefficient is [00040] and minimum divisions are [D=2],
[range
to detect stability]= 0.5D × 40 × 2
= 40D
7−7−2. Time to detect stability
Execute the setting of time to detect stability with the function F−65.
It can be set the range of [00000] 〜 [00099] every [0.1]s.
OFF when set the [00000]. Default is set as [00005].
[time
to detect stability]= 0.1s × [time to detect stability coefficient]