7−8. Various kinds of functions related with the display
7−8−1. Selection of display rate
The instrument can select the display times from the “4 times/s”, “20 times/s”, “50 times/s” and
“100 times/s”. This selection is available in the Function mode. (Related function F−03).
As for the default, 20 times/s has selected.
7−8−2. Selection of decimal point display position
The instrument can display the decimal point at the “Load display section” of the instrument.
The selection of display can be made in the Function mode. (Related function F−01) As for the
default, “No decimal point display” has selected.
7−8−3. Load display range
The load display range for the instrument is fixed from the −10 % to 110 % of the maximum
display value at the time of setting during calibration.
When less than the range, “−OL” displays and over the range “OL” displays. For example,
when the maximum display value is “1 000”, the load display range will be from −100 to 1 100.
Besides, when under −100, the “−OL” displays, and over 1 100, the “OL” will display.
7−9. Selection the target for HOLD
The instrument can select the target for HOLD function among “load display”, “comparator S0
open collector output, LED display”, “comparator S1, S2 open collector output, LED display”,
“Analog output” and “Optional BCD output”.
This selection can be made in the Function mode(Related function F−10).
As for the default, “All is selected.” Layout of setting for the F−10 are as follows:
10
0
digit:Load display
10
1
digit:Comparator S0 open collector output, LED display
10
2
digit:Comparator S1, S2 open collector output, LED display
10
3
digit:Optional analog output
10
4
digit:Optional BCD output
※With the ”0” setting, out of the target, and with “1” setting, target of HOLD.
7−10. Change of bridge power supply voltage
The instrument can select the bridge power supply from “10 V”, “5 V” and “2.5 V”. This selection
can be made in the Function mode(Related function F−12). As for the default, “10 V” has selected.
● When the bridge power supply voltage is changed, make calibration
again.