123
D F−31 Setting the target of comparator
Initial value=00000 0=Gross weight
1=Net weight
10
0
digit:Comparator S0
10
1
digit:Comparator S1
10
2
digit:Comparator S2
10
3
digit:Comparator S3
10
4
digit:Comparator S4
※The comparator S0 is effective in the setting 00007 by F−33.
D F−32 Setting the direction of comparator
Initial value=00000 0=or more
1=or less
10
0
digit:Comparator S0
10
1
digit:Comparator S1
10
2
digit:Comparator S2
10
3
digit:Comparator S3
10
4
digit:Comparator S4
※The comparator S0 is effective in the setting 00007 by F−33.
D F−33 Setting applicable condition for the comparator S0
Initial value=00000 00000=ON when load display value≧the maximum display value
00001=ON when both S1 and S2 open collector output are OFF.
00002=ON when both S1 and S3 open collector output are OFF.
00003=ON when both S1 and S4 open collector output are OFF.
00004=ON when both S2 and S3 open collector output are OFF.
00005=ON when both S2 and S4 open collector output are OFF.
00006=ON when both S3 and S4 open collector output are OFF.
00007=Operation whichever “open collector ON over the set
value” or “open collector ON less than the set value”
00008=Interlocked with HOLD LED
00009=Interlocked with A/Z LED
00010=Interlocked with LOCK LED
00011=ON when the detection of stability
D F−34 Setting the condition of Hysteresis operation for comparator
Initial value=00000 00000=ON delay
00001=OFF delay
D F−35 Hysteresis data width OFF
Initial value=00000 00000=Hysteresis data width OFF
Setting rage:00000〜00099
Unit:1D
Data width of 99D at the setting of “00099”.